FDA device recall Z-1185-2016 was initiated by Stryker Medical Division Of Stryker Corporation on Feb 5, 2016 and is designated Class II. Reason for recall: The hydraulic jack assemblies (PN: 0753-002-101, 0753-002-070, 0753-002-001, and 0753-002-001) were manufactured using the incorrect oil. The oil used does not contain the same wear resistance… The recall status is terminated (terminated Sep 6, 2016). Affected quantity: 36.

Recall details

Product Type
Devices
Report Date
Mar 23, 2016
Initiation Date
Feb 5, 2016
Termination Date
Sep 6, 2016
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
36

Product description

Transport Stretcher, and Head and Neck Surgery Stretcher Non-powered stretcher

Reason for recall

The hydraulic jack assemblies (PN: 0753-002-101, 0753-002-070, 0753-002-001, and 0753-002-001) were manufactured using the incorrect oil. The oil used does not contain the same wear resistance additives as the specified oils and therefore its use increases the potential for wear which may cause a leak event or jack drift. Testing was done to evaluate the jacks with incorrect oil and leak events we

Distribution pattern

Distributed in DC and the states of AL, CA, FL, MN, MT, NC, NE, NY, PA, RI, TN, TX, UT, and WV.

Code information

The hydraulic jack assemblies (Part numbers: 0753-002-101, 0753-002-070, 0753-002-001, and 0753-002-001) used as service parts on Stretchers assembled from 8/28/15 - 8/29/15

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1185-2016
openFDA file date
Nov 16, 2026
Loaded by RegDataLab
Nov 17, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Source: openFDA record for Z-1185-2016. Report a data issue: hello@regdatalab.com.