Z-1185-2026 Class II Ongoing
Recall Details
- Product Type
- Devices
- Report Date
- February 4, 2026
- Initiation Date
- December 12, 2025
- Termination Date
- N/A
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 499 units
Product Description
Conductivity Standard Solution Catalog/Model Numbers: 10001, 10011, 10021, 10032, 10040, 10040c, and 10041 Conductivity Standard Solutions are a secondary standard solution used for the calibration of conductivity cells together with conductivity meters
Reason for Recall
due to a defect associated with the conductivity values of the solution used to calibrate meters and result in the reporting of incorrect conductivity values for dialysate.
Distribution Pattern
U.S. Nationwide distribution in the states of AL, AR, AZ, CA, DE, FL, GA. IL, IN, KS, MI, MO, NC, ND, NJ, NV, NY, OH, OK, OR, PA, SC, TN TX, UT, VA, and WA.
Code Information
Catalog/Model Numbers/ Lot Numbers / UDI codes: 10001 14250616-1019106 (01) 0 0850016 97600 0 (17) 251030 (10) SHOWCASE 10011 14250616-1019107 (01) 0 0850016 97609 3 (17) 251030 (10) SHOWCASE 10040c 50250624-1019120 (01) 0 0850016 97636 9 (17) 251030 (10) SHOWCASE 10040 50250624-1019119 (01) 0 0850016 97635 2 (17) 251030 (10) SHOWCASE 10032 14250616-1019108 (01) 0 0850016 97631 1 (17) 251030 (10) SHOWCASE 10041 50250522-1018643 (01) 0 0850016 97637 6 (17) 251030 (10) SHOWCASE 10041 50250624-1019121 (01) 0 0850016 97637 6 (17) 251030 (10) SHOWCASE 10021 15F250603-1018834 (01) 0 0850016 97619 2 (17) 251030 (10) SHOWCASE