Z-1198-2025 Class II Ongoing

Recalled by Augustine Temperature Management, LLC — Minnetonka, MN

Recall Details

Product Type
Devices
Report Date
February 26, 2025
Initiation Date
December 16, 2024
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
1 unit

Product Description

Hot Dog PATIENT WARMING, Underbody Warming Mattress + Return Electrode, 127 cm (50"), REF US502, to prevent or treat hypothermia

Reason for Recall

There is over-molding that stabilizes the connection between the cable and cable-entry on the mattress. It turns out that after about a year of use, the connector cable on the mattress can experience the same type of deflection that caused the intermittent connection with the cables connecting to the ESU.

Distribution Pattern

US Nationwide distribution in the states of TX, AL, NV; AU, NZ.

Code Information

UDI-DI UDI-DI 00855913001824, Serial Numbers: 1304283