Z-1210-2013 Class II Terminated
FDA device recall Z-1210-2013 was initiated by Stryker Howmedica Osteonics Corp. on Mar 28, 2013 and is designated Class II. Reason for recall: Stryker Orthopaedics has received a report indicating that the scale of the AxSOS Calibrated Drill Bit 4.3,,X263mm, flat , manufactured by Stryker Osteosynthesis, is incorrect. The incorrect scale… The recall status is terminated (terminated Nov 17, 2015). Affected quantity: 107 units.
Recall details
- Product Type
- Devices
- Report Date
- May 8, 2013
- Initiation Date
- Mar 28, 2013
- Termination Date
- Nov 17, 2015
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 107 units
Product description
Stryker Howmedica Osteonics Corp. AxSOS System Instrument. The AxSOS Locking Plate System is designed to treat periarticular or intraarticular fractures of the Distal Femur, Proximal Humerus, Proximal Tibia, and the Distal Tibia. Per the operative technique, " A 4.3 mm Drill (REF 702743) is then used to drill the core hole for the locking screw. Using a fluoroscopy, check the correct depth of the drill, and measure the length of the screw. The Drill Sleeve should now be removed, and the correct length 5.0 mm Locking Screw is inserted using the Screwdriver T20 and Screw Holding Sleeve."
Reason for recall
Stryker Orthopaedics has received a report indicating that the scale of the AxSOS Calibrated Drill Bit 4.3,,X263mm, flat , manufactured by Stryker Osteosynthesis, is incorrect. The incorrect scale could result in the surgeon selecting a screw that is 10mm longer than intended.
Distribution pattern
Nationwide Distribution including the states of AL, AZ, IL, KY, LA, NJ, NC, FL, CA, NM, MO, TN, VA, CO, MA, TX, IN and MT.
Code information