Z-1216-2017 Class III Terminated
FDA device recall Z-1216-2017 was initiated by Siemens Healthcare Diagnostics, Inc. on May 5, 2016 and is designated Class III. Reason for recall: Complaints of positive shift of patient and QC results when the ALTI assay is calibrated with the Dimension Vista Enzyme 2 calibrator. Internal investigation has confirmed that patient samples and QC… The recall status is terminated (terminated Feb 2, 2018). Affected quantity: 13,147 units.
Recall details
- Product Type
- Devices
- Report Date
- Feb 22, 2017
- Initiation Date
- May 5, 2016
- Termination Date
- Feb 2, 2018
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 13,147 units
Product description
The Enzyme 2 Calibrator is an in vitro diagnostic product for the calibration of Alanine Aminotransferase (ALTI and ALTI) and Aspartate Aminotransferase (AST) methods on the Dimension Vista System. For prescription use only.
Reason for recall
Complaints of positive shift of patient and QC results when the ALTI assay is calibrated with the Dimension Vista Enzyme 2 calibrator. Internal investigation has confirmed that patient samples and QC material demonstrate a shift of up to 10% in comparison to results obtained using earlier calibrator lots.
Distribution pattern
Distributed throughout the United States
Code information
Lot Number: 4MD048, 5DD026, 5ED071, 5GD004, 5JD058, 5LD061