Z-1225-2018 Class II Terminated
Recalled by Zimmer Biomet, Inc.
— Warsaw, IN
FDA device recall Z-1225-2018 was initiated by Zimmer Biomet, Inc. on Dec 20, 2017 and is designated Class II. Reason for recall: Zimmer Biomet is conducting a medical device recall for the ACE Trochanteric Nail (ATN) system after a Device Performance Review (DPR) failed clinical analysis for survivorship pursuant to the… The recall status is terminated (terminated Apr 22, 2020). Affected quantity: 3454.
Recall details
- Product Type
- Devices
- Report Date
- Apr 4, 2018
- Initiation Date
- Dec 20, 2017
- Termination Date
- Apr 22, 2020
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 3454
Product description
ACE Trochanteric Nail System ACE TROCH NAIL 125 DEG. X 11MM ACE TROCH NAIL 130 DEG. X 11MM ACE TROCH NAIL 135 DEG. X 11MM ACE TROCH NAIL 125 DEG. X 13MM ACE TROCH NAIL 130 DEG. X 13MM ACE TROCH NAIL 135 DEG. X 13MM The Trochanteric Nail System is intended to treat stable and unstable proximal fractures of the femur including pertrochanteric fractures, intertrochanteric fractures, high subtrochanteric fractures and combinations of these fractures. The Trochanteric Long Nail System is additionally indicated to treat pertrochanteric fractures associated with shaft fractures, pathologic fractures in osteoporotic bone of the trochanteric and diaphyseal areas, long subtrochanteric fractures, ipsilateral femoral fractures, proximal or distal nonunions and malunions and revision procedures.
Reason for recall
Zimmer Biomet is conducting a medical device recall for the ACE Trochanteric Nail (ATN) system after a Device Performance Review (DPR) failed clinical analysis for survivorship pursuant to the acceptance criteria for the DPR protocol for the ATN system.
Distribution pattern
USA (nationwide)
Code information
903011225 903011230 903011235 903013225 903013230 903013235
Source
- Dataset
- openFDA device recall enforcement reports
- FDA record
- This recall in openFDA (JSON)
- Enforcement Report
- Search FDA's Enforcement Report for Z-1225-2018
- openFDA file date
- Nov 24, 2026
- Loaded by RegDataLab
- Nov 25, 2026
These dates say when openFDA published the data file and when RegDataLab last loaded it.
Need the full FDA record for Zimmer Biomet, Inc.? Ask Claude or ChatGPT with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. The API key is free: no card.
Get a free API key