Z-1230-2020 Class II Terminated

Recalled by Siemens Healthcare Diagnostics, Inc. — Tarrytown, NY

FDA device recall Z-1230-2020 was initiated by Siemens Healthcare Diagnostics, Inc. on Dec 18, 2019 and is designated Class II. Reason for recall: There is a potential for falsely depressed creatinine results for patients on phenindione therapy when using the enzymatic methodology. Interference has not been observed with the creatinine Jaffe… The recall status is terminated (terminated Jun 25, 2020). Affected quantity: 1,015 units.

Recall details

Product Type
Devices
Report Date
Feb 26, 2020
Initiation Date
Dec 18, 2019
Termination Date
Jun 25, 2020
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
1,015 units

Product description

ADVIA Chemistry Enzymatic Creatinine (ECRE_2) reagent, SMN 10335869

Reason for recall

There is a potential for falsely depressed creatinine results for patients on phenindione therapy when using the enzymatic methodology. Interference has not been observed with the creatinine Jaffe methodology.

Distribution pattern

US Nationwide

Code information

UDI Number: 00630414526690 Lots: All lots

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1230-2020
openFDA file date
Nov 25, 2026
Loaded by RegDataLab
Nov 26, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Siemens Healthcare Diagnostics, Inc.? Ask Claude with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. Free to start, no card.

Question to ask in Claude

Use RegDataLab to give me an FDA risk summary for Siemens Healthcare Diagnostics, Inc.: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.

Ask Claude about Siemens Healthcare Diagnostics, Inc. →

Source: openFDA record for Z-1230-2020. Report a data issue: hello@regdatalab.com.