Z-1232-2023 Class II Ongoing

Recalled by Datascope Corp. — Mahwah, NJ

FDA device recall Z-1232-2023 was initiated by Datascope Corp. on Feb 7, 2023 and is designated Class II. Reason for recall: Firm has received reports of damaged, worn, or torn O-rings on the Cardiosave Pump Console quick disconnect fitting, which results in helium tank leaks. If the helium supply is depleted, therapy will… The recall status is ongoing. Affected quantity: 47 units (44 US, 3 OUS).

Recall details

Product Type
Devices
Report Date
Mar 15, 2023
Initiation Date
Feb 7, 2023
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
47 units (44 US, 3 OUS)

Product description

Cardiosave Rescue. Model Number: 0998-00-0800-83.

Reason for recall

Firm has received reports of damaged, worn, or torn O-rings on the Cardiosave Pump Console quick disconnect fitting, which results in helium tank leaks. If the helium supply is depleted, therapy will be interrupted.

Distribution pattern

Nationwide US distribution. Worldwide international distribution.

Code information

Model Number: 0998-00-0800-83. UDI-DI: 10607567108407. All serial numbers.

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1232-2023
openFDA file date
Nov 10, 2026
Loaded by RegDataLab
Nov 11, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Source: openFDA record for Z-1232-2023. Report a data issue: hello@regdatalab.com.