Z-1237-2019 Class II Ongoing

Recalled by PerkinElmer Life and Analytical Sciences, Wallac, OY — Turku, N/A

Recall Details

Product Type
Devices
Report Date
May 8, 2019
Initiation Date
January 10, 2019
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
31 kits (10 in US, 21 in ROW)

Product Description

DELFIA¿ hCG Kit, In Vitro Diagnostic for the determination of human gonadotropin (hCG) Product Code: A007-101 - Product Usage: Quantitative determination of human gonadotrophin (hCG) in serum.

Reason for Recall

hCG analyte concentrations may be decreased up to 15- 20% at the lower measuring range of hCG concentrations.

Distribution Pattern

Worldwide distribution - US Nationwide distribution. and countries of Austria, Bolivia, UK, Switzerland,Thailand, Finland.

Code Information

Kit Lot Numbers: 664891, 666471, 666604 Pack Lot Numbers: 1066489101, 1066647101,1066660401 UDI: (01)06438147009176(17)190430(10)664891 (01)06438147009176(17)190430(10)666471 (01)06438147009176(17)190430(10)666604