Z-1241-2016 Class II Terminated

Recalled by Siemens Healthcare Diagnostics, Inc — East Walpole, MA

FDA device recall Z-1241-2016 was initiated by Siemens Healthcare Diagnostics, Inc on Jan 13, 2016 and is designated Class II. Reason for recall: The drug fulvestrant (Faslodex¿) may cause falsely elevated estradiol results. The recall status is terminated (terminated Jan 8, 2018). Affected quantity: 1945 units.

Recall details

Product Type
Devices
Report Date
Mar 30, 2016
Initiation Date
Jan 13, 2016
Termination Date
Jan 8, 2018
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
1945 units

Product description

Dimension Vista LOCI Estradiol Test Code E2 Catalog Number: K6463 SMN: 10489099 Product Usage: The E2 method is an in vitro diagnostic test for the quantitative measurement of estradiol in human serum and plasma on the Dimension Vista¿System. Estradiol measurements are used in the diagnosis and treatment of various hormonal sexual disorders and in assessing placental function in complicated pregnancy.

Reason for recall

The drug fulvestrant (Faslodex¿) may cause falsely elevated estradiol results.

Distribution pattern

Worldwide Distribution  US Nationwide , AR, AU, AU, BD, BO, BR, CA, CH, CL, CN, CO, DO, EC, EG, HK, ID, IE, IL, IN, JP, KR, LK, LY, MX, MY, NI, NP, NZ, PE, PH, PY, SA, SG, SV, TH, TW, UY, VE, VN, and ZA.

Code information

All lots

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1241-2016
openFDA file date
Nov 27, 2026
Loaded by RegDataLab
Nov 28, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Siemens Healthcare Diagnostics, Inc? Ask Claude with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. Free to start, no card.

Question to ask in Claude

Use RegDataLab to give me an FDA risk summary for Siemens Healthcare Diagnostics, Inc: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.

Ask Claude about Siemens Healthcare Diagnostics, Inc →

Source: openFDA record for Z-1241-2016. Report a data issue: hello@regdatalab.com.