Z-1246-2025 Class I Ongoing

Recalled by Micro Therapeutics, Inc. — Irvine, CA

FDA device recall Z-1246-2025 was initiated by Micro Therapeutics, Inc. on Jan 30, 2025 and is designated Class I. Reason for recall: Use of embolization device with diameters greater-than-or-equal-to 4mm may lead to incomplete wall apposition and/or braid deformation (fish-mouthing/braid narrowing/braid collapse) noted during… The recall status is ongoing. Affected quantity: 7820.

Recall details

Product Type
Devices
Report Date
Mar 12, 2025
Initiation Date
Jan 30, 2025
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
7820

Product description

Medtronic Pipeline Vantage Embolization Device with Shield Technology, REF: PED3-021-350-16, PED3-021-250-14, PED3-021-350-20, PED3-021-250-20, PED3-021-300-16, PED3-021-300-20, PED3-021-275-12, PED3-021-275-16, PED3-021-325-12, PED3-021-325-14, PED3-021-325-16, PED3-021-325-20, PED3-021-250-12, PED3-021-300-14, PED3-021-250-16, PED3-021-300-12, PED3-021-275-14, PED3-021-350-14, PED3-021-350-25, PED3-021-350-12, PED3-021-250-10, PED3-021-275-20, PED3-021-350-18, PED3-021-325-18, PED3-021-275-18, PED3-021-250-18, PED3-021-300-18; and Instructions for Use, Part: M056989CDOC2

Reason for recall

Use of embolization device with diameters greater-than-or-equal-to 4mm may lead to incomplete wall apposition and/or braid deformation (fish-mouthing/braid narrowing/braid collapse) noted during procedure and post-procedure, typically noted at 6-12 month imaging follow-up, potentially can lead to thrombosis and/or serious adverse events. Higher risk in females less-than-or-equal-to 45 years of age

Distribution pattern

Worldwide distribution: US (nationwide): PR, OK, CA, MD, FL, NY, WA, MI, AL, NV, MN, NJ, IL, OH, AK, CT, TN, MO, TX, ID, AZ, WI, PA, OR, SD, VA, CO, GA, KY, IN, AR, MA, LA, NE, MT, ME, SC, DC, NC, ND, NM, IA, UT, KS, WV, NH, HI, RI. And OUS (foreign) countries of: Algeria, Argentina, Armenia, Australia, Austria, Bahrain, Bangladesh, Belgium, Brazil, Bulgaria, Canada, Canary Islands, Chile, Colombia, Costa Rica, Croatia, Cyprus, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, Finland, France, Germany, Greece, Hong Kong, Hungary, Iceland, India, Ireland, Israel, Italy, Jordan, Kazakhstan, Kuwait, Latvia, Lebanon, Lithuania, Luxembourg, Malaysia, Martinique, Mexico, Republic Of Moldova, Montenegro, Morocco, Netherlands, New Zealand, North Macedonia, Northern Ireland, Norway, Panama, Peru, Philippines, Poland, Portugal, Qatar, Reunion, Romania, Russia, Saudi Arabia, Serbia, Singapore, Slovakia, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Thailand, Tunisia, Turkey, Ukraine, United Arab Emirates, United Kingdom, Uruguay, Vietnam

Code information

Instructions for Use, Version: Rev. A (01/2024). REF/UDI-DI/Lot: PED3-021-350-16/00847536030701, 00847536035195, 00847536036529/B317460, B327449, B335184, B340948, B349032, B389162, B392918, B424454, B430015, B495821, B497938, B547396, B550307, B551615, B559033, B603377, B623564, B656863, B656887, B658904, B692346, B715057, B715058, B715541, B726894, B735821, B735823, B756029, B756030, B760210, B760211, B768202, B768387, B769453, B770633, B807494; PED3-021-250-14/00847536030411, 00847536034907, 00847536036239/B317936 B332767, B333492, B333536, B334545, B376967, B388041, B391185, B399354, B416713, B431395, B452416, B457963, B467691, B476239, B564173, B564970, B574472, B580923, B629372, B630998, B670305, B673035, B699576, B707568, B709193, B736698, B736893, B739090, B739091, B795110, B796586, B800827, B801770; PED3-021-350-20/00847536030725, 00847536035218, 00847536036543/B319436, B336571, B379993, B383427, B389708, B400103, B400599, B403612, B407062, B415211, B446722, B483987, B490972, B492403, B494049, B552487, B559741, B591782, B591864, B592074, B622656, B626023, B648708, B684905, B715692, B719565, B719567, B719659, B726923, B728926, B735667, B756330, B765053, B768231, B770237, B770248, B792630; PED3-021-250-20/00847536030442, 00847536034938, 00847536036260/B321465, B337378, B360749, B371070, B412868, B467843, B477487, B488713, B548244, B557602, B644034, B673388, B679822, B684208, B686727, B695486, B698571, B703035, B708080, B709204, B711078, B746067, B757056, B757060, B761030, B771498, B775215, B789892, B790055; PED3-021-300-16/00847536030541, 00847536035034, 00847536036369/B321725, B338497, B405720, B418200, B570723, B574473, B603375, B604477, B656277, B658580, B659806, B666425, B681289, B691752, B699606, B703209, B707982, B711365, B725679, B736659, B738114, B769551, B769582, B798864; PED3-021-300-20/00847536030565, 00847536035058, 00847536036383/B322263, B336066, B347327, B376257, B405892, B414535, B458672, B458776, B469040, B490920, B495292, B563197, B607500, B607501, B635214, B636992, B656299, B656355, B657960, B668529, B678482, B680045, B681190, B712110, B712111, B718522, B720923, B725457, B732895, B735956, B737156, B737367, B748272, B748683, B749541, B775230, B778774, B780044; PED3-021-275-12/00847536030466, 00847536034952, 00847536036284/B323044, B383016, B391275, B399486, B454385, B468406, B623580, B623583, B624424, B657776, B666498, B669520, B670179, B670226, B710976, B743820, B743822, B754659, B762034, B768199, B798851; PED3-021-275-16/00847536030480, 00847536034976, 00847536036307/B323046, B361987, B406342, B414102, B458076, B468915, B477868, B558538, B562688, B563304, B603078, B654235, B656399, B657059, B666486, B669610,B670343, B677246, B699308, B736874, B743098, B755194, B755196, B755673, B769461, B771317, B775167, B775168; PED3-021-325-12/00847536030602, 00847536035096, 00847536036420/B324297, B407597, B418843, B494876, B561787, B592073, B623579, B629046, B676507, B678997, B716220, B741395, B751748, B787015, B787017; PED3-021-325-14/00847536030619, 00847536035102, 00847536036437/B324809, B347993, B362685, B459598, B558177, B563209, B575964, B580901, B628946, B675724, B678948, B681904, B714342, B714344, B733352, B733353, B769593, B790478; PED3-021-325-16/00847536030626, 00847536035119, 00847536036444/B324954, B339033, B371691, B418451, B460439, B495941, B603378, B604475, B623578, B624422, B660509, B664402, B666503, B667510, B667920, B668523, B680717, B681747, B733912, B739722, B739723, B739724, B746618, B794191, B794192; PED3-021-325-20/00847536030640, 00847536035133, 00847536036468/B325613, B371743, B418905, B421650, B463895, B487451, B495368, B591952, B591955, B644630, B644769, B644809, B657857, B659135, B659138, B659243, B684852, B716733, B716735, B741305, B741316, B741321, B787131; PED3-021-250-12/00847536030404, 00847536034891, 00847536036222/B326236, B335270, B387931, B390988, B394283, B421249, B477292, B497244, B578963, B664457, B666521, B667945, B669726, B696189, B719158, B724199, B724200, B724764, B741102, B746521, B764797, B764798, B787764, B787765, B787766, B793905; PED3-021-300-14/00847536030534, 00847536035027, 00847536036352/B327828, B335957, B346789, B393640, B399293, B403772, B456576, B547554, B568649, B575299, B631125, B631126, B649258, B650123, B652886, B657773, B683152, B711496, B712157, B712158, B732804, B732807, B754755, B754756, B754757, B760951, B760953, B768142, B769947, B789256; PED3-021-250-16/00847536030428, 00847536034914, 00847536036246/B328095, B416673, B452284, B478579, B494048, B565441, B573470, B575302, B628868, B628869, B630965, B656206, B666535, B667323, B668743, B739139, B739405, B739408, B754155, B754886, B754888, B755651, B794406, B794408, B794426; PED3-021-300-12/00847536030527, 00847536035010, 00847536036345/B338969, B345449, B346121, B361988, B405158, B418358, B454923, B551608, B561782, B562698, B563303, B575894, B677108, B677319, B699289, B703245, B718483, B718491, B739905, B744728, B760716, B766588; PED3-021-275-14/00847536030473, 00847536034969, 00847536036291/B343142, B361366, B379727, B429213, B448933, B452378, B461124, B601345, B607550, B624425, B626121, B678023, B680119, B711192, B713287, B724745, B744169, B744175, B744176, B770971, B770988; PED3-021-350-14/00847536030695, 00847536035188, 00847536036512/B349031, B372357, B388981, B391924, B399969, B421758, B445867, B491603, B492310, B497318, B596593, B629649, B630549, B631109, B670495, B678371, B679957, B713023, B713796, B740630, B740631, B742526, B743935, B753618; PED3-021-350-25/00847536030732, 00847536035225, 00847536036550/B364361, B393438, B403082, B419021, B422233, B430531, B437842, B447224, B488266, B616574, B624414, B629444, B629660, B672265, B672892, B673364, B717131, B717132, B724110, B733587, B733590, B741976, B761681, B768780, B771211, B772034, B773456, B773457; PED3-021-350-12/00847536030688, 00847536035171, 00847536036505/B368529, B406425, B419491, B490325, B496666, B497114, B611908, B615889, B616569, B643508, B643509, B644619, B657109, B658427, B659171, B690687, B715610, B715612, B716621, B742378, B746455, B787264; PED3-021-250-10/00847536030398, 00847536036215, 00847536034884/B387475, B390553, B427939, B490270, B575123, B578342, B628755, B654233, B666402, B670376, B693553, B707945, B711113, B740699, B764024, B764025, B766116, B769419, B797931, B801945; PED3-021-275-20/00847536030503, 00847536034990, 00847536036321/B391792, B398628, B418296, B452832, B461730, B497830, B548241, B567772, B655966, B656945, B666505, B671729, B677211, B679094, B680758, B693403, B699179, B711385, B725638, B725639, B736012, B736612, B737225, B751998, B756411, B797986; PED3-021-350-18/00847536036536/B798906; PED3-021-325-18/00847536036451/B800239; PED3-021-275-18/00847536036314/B805162; PED3-021-250-18/00847536036253/B805488; PED3-021-300-18/00847536036376/B807417

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1246-2025
openFDA file date
Nov 25, 2026
Loaded by RegDataLab
Nov 26, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Micro Therapeutics, Inc.? Ask Claude with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. Free to start, no card.

Question to ask in Claude

Use RegDataLab to give me an FDA risk summary for Micro Therapeutics, Inc.: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.

Ask Claude about Micro Therapeutics, Inc. →

Source: openFDA record for Z-1246-2025. Report a data issue: hello@regdatalab.com.