Z-1247-2014 Class II Terminated

Recalled by Stryker Instruments Div. Of Stryker Corporation — Portage, MI

FDA device recall Z-1247-2014 was initiated by Stryker Instruments Div. Of Stryker Corporation on Mar 18, 2014 and is designated Class II. Reason for recall: Stryker Instruments is voluntarily recalling the single use 2.5MM Aggressive Blade because the line on the affected part (P/N 5290-928-000, Lot #13280CG2) was a straight line without a number scale or… The recall status is terminated (terminated May 15, 2014). Affected quantity: 90 units.

Recall details

Product Type
Devices
Report Date
Mar 26, 2014
Initiation Date
Mar 18, 2014
Termination Date
May 15, 2014
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
90 units

Product description

2.5MM Aggressive Blade Known as a Sinuscopic Shaver Blade, this device is composed of a metal cutter and plastic drive shaft that goes into a metal housing supported by a plastic hub. ENT Precision Cutter, Aggressive, ESSxlHUMMER 4, Rx Only It is used for ENT procedures and functions by means of rotation when adapted to a handpiece and console. Through the rotation, the necessary tissue is resected as required by the user. The blade housing has an etching line guide that starts with a straight line and along with a scale from 1cm to 7cm in 1cm increments that serves the doctor as a guide by indicating how far the device is being introduced into the surgical area.

Reason for recall

Stryker Instruments is voluntarily recalling the single use 2.5MM Aggressive Blade because the line on the affected part (P/N 5290-928-000, Lot #13280CG2) was a straight line without a number scale or logo. The line required for this part number includes a number scale that goes from 1cm  7cm.

Distribution pattern

US Distribution including the states of NV, CA, WA, TN, OR, SD, ID, IN and CO., and Internationally to Malaysia

Code information

P/N 5290-628-000, Lot #13280CG2

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1247-2014
openFDA file date
Nov 10, 2026
Loaded by RegDataLab
Nov 11, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Source: openFDA record for Z-1247-2014. Report a data issue: hello@regdatalab.com.