Z-1249-2022 Class II Ongoing

Recalled by Getinge Usa Sales Inc — Wayne, NJ

FDA device recall Z-1249-2022 was initiated by Getinge Usa Sales Inc on Apr 13, 2022 and is designated Class II. Reason for recall: Verification of installation not documented or may be incomplete, could result in harm to the user such as skin burns, smoke inhalation, electric shock, or a crush or pinch injury or slip and fall… The recall status is ongoing. Affected quantity: 11 units.

Recall Details

Product Type
Devices
Report Date
Jun 15, 2022
Initiation Date
Apr 13, 2022
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
11 units

Product Description

Getinge CM320 WUWD Series Washer Disinfector-For use as a multi-chamber washer-disinfector for the washing, disinfecting and drying surgical instruments, Model: CM320 WUWD

Reason for Recall

Verification of installation not documented or may be incomplete, could result in harm to the user such as skin burns, smoke inhalation, electric shock, or a crush or pinch injury or slip and fall injury. May lead to a patient procedural delay and/or infection

Distribution Pattern

US Nationwide distribution in the states of MD, MN, ND, NE, NY, PA, VA.

Code Information

UDI-DI: 07340153700192 Serial Numbers: W50026718 W50029991 WAA081312 W50039234 WAA062340 WAA062341 WAA070619 WAA085063 WAA089556 WAA093936 WAA094247

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1249-2022
openFDA file date
Oct 23, 2026
Loaded by RegDataLab
Oct 26, 2026