Z-1253-2019 Class III Terminated

Recalled by Carolina Liquid Chemistries — Greensboro, NC

FDA device recall Z-1253-2019 was initiated by Carolina Liquid Chemistries on January 16, 2019 and is designated Class III. Reason for recall: Supplier testing showed that the absorbance rate of the Ethyl Alcohol Assay did not meet internal quality control specifications. The defective lots will not function toward the end of shelf life. The recall status is terminated (terminated February 2, 2021). Affected quantity: 20 kits.

Recall Details

Product Type
Devices
Report Date
May 8, 2019
Initiation Date
January 16, 2019
Termination Date
February 2, 2021
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
20 kits

Product Description

CAROLINA(R) LIQUID CHEMISTRIES CORP. ETHA RGT. KIT, IVD, REF ST321

Reason for Recall

Supplier testing showed that the absorbance rate of the Ethyl Alcohol Assay did not meet internal quality control specifications. The defective lots will not function toward the end of shelf life.

Distribution Pattern

NY. NH, ME, NC, TX, AR, SC, CT, IL, CA, NJ

Code Information

Lot 11687