Z-1267-2024 Class I Ongoing

Recalled by ARROW INTERNATIONAL Inc. — Morrisville, NC

FDA device recall Z-1267-2024 was initiated by ARROW INTERNATIONAL Inc. on Feb 12, 2024 and is designated Class I. Reason for recall: Teleflex is initiating this voluntary recall for the above-mentioned products due to reports received indicating a potentially defective component. The complaints received relate to resistance of the… The recall status is ongoing. Affected quantity: 1,160,470 devices.

Recall Details

Product Type
Devices
Report Date
Mar 27, 2024
Initiation Date
Feb 12, 2024
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
1,160,470 devices

Product Description

ARROW QuickFlash Radial Artery Catherization (wire, guide, catheter), Part Numbers: a) REF AK-04220; b) REF ASK-04220-FMH1; c) REF ASK-04220-HHC; d) REF ASK-04220-KSP; e) REF ASK-04220-PSU; f) REF ASK-04220-UCL1; g) REF ASK-04500-AH; h) REF ASK-04500-HF-S; i) REF NA-04220-S1A; j) REF NA-04220-X1A; k) REF RA-04220; l) REF RA-04220-W

Reason for Recall

Teleflex is initiating this voluntary recall for the above-mentioned products due to reports received indicating a potentially defective component. The complaints received relate to resistance of the guidewire handle/chamber during use. The possible immediate health consequences of the component issue are arterial vasospasm and vessel injury arising from multiple arterial punctures with repeated attempts.

Distribution Pattern

Worldwide distribution: US (nationwide) and OUS (foreign) to countries of: AE, AR, AT, AU, BE, BH, BR, CA, CN, CO, CY, DE, EC, EE, ES, FR, GR, HK, HR, IT, KW, LB, LU, MX, MY, NL, NZ, PH, PL, QA, RU, SA, SE, SG, SI, TH, TW, and UK.

Code Information

a) REF AK-04220, UDI/DI 20801902190943, Lot Numbers: 13F22E0436, 13F22H0360, 33F23E0375, 33F23K0032, 33F23J0443, 33F23L0862; UDI/DI 30801902121616, Lot Numbers: 13F22L0105, 13F23A0651, 33F23C0100, 33F23F0347, 33F23E0501, 33F23G0141; b) REF ASK-04220-FMH1, UDI/DI 20801902111665, Lot Numbers: 33F23B0525, 33F23H0100; c) REF ASK-04220-HHC, UDI/DI 20801902190851, Lot Numbers: 33F23F0067, 33F23K0278, 33F23G0478, 33F23H0101; d) REF ASK-04220-KSP, UDI/DI 20801902156406, Lot Numbers: 33F23E0709, 33F23H0102, 33F23H0918, 33F23J0607; e) REF ASK-04220-PSU, UDI/DI 40801902121446, Lot Numbers: 33F23F0331, 33F23G0663; f) REF ASK-04220-UCL1, UDI/DI 20801902159124, Lot Numbers: 33F23F0801, 33F23C0726, 33F23H0320; g) REF ASK-04500-AH, UDI/DI 20801902114048, Lot Numbers: 33F23E0731, 33F23H0103, 33F23H0395; h) REF ASK-04500-HF-S, UDI/DI 40801902118002, Lot Numbers: 33F23F0202, 33F23H0104, 33F23H0396; i) REF NA-04220-S1A, UDI/DI 20801902139416, Lot Numbers: 33F23J0671; j) REF NA-04220-X1A, UDI/DI 20801902136606, Lot Numbers: 33F23F0704; k) REF RA-04220, UDI/DI 20801902180678, Lot Numbers: 14F23B0041; UDI/DI 30801902193309, Lot Numbers: 14F21L0189, 14F21L0190, 14F21L0250, 14F21L0251, 14F21L0252, 14F21L0253, 14F21M0130, 14F21M0131, 14F21M0205, 14F21M0206, 14F22A0044, 14F22A0123, 14F22A0124, 14F22A0143, 14F22A0144, 14F22B0163, 14F22B0164, 14F22B0175, 14F22C0017, 14F22C0018, 14F22C0019, 14F22C0020, 14F22C0021, 14F22C0022, 14F22C0216, 14F22C0284, 14F22C0346, 14F22C0347, 14F22D0023, 14F22D0024, 14F22D0110, 14F22D0217, 14F22D0291, 14S22D0218, 14F22E0016, 14F22E0173, 14F22E0215, 14F22F0003, 14F22F0062, 14F22F0208, 14F22F0209, 14F22F0210, 14F22F0212, 14F22F0335, 14F22F0336, 14F22G0297, 14F22G0298, 14F22G0325, 14F22G0326, 14F22H0078, 14F22H0079, 14F22H0250, 14F22J0078, 14F22J0079, 14F22J0080, 14F22J0081, 14F22J0082, 14F22J0083, 14F22J0257; UDI/DI 30801902210631, Lot Numbers: 14F22K0182, 14F22K0183, 14F22K0184, 14F22K0185, 14F22K0223, 14F22M0032, 14F22M0033, 14F22M0034, 14F22M0035, 14F22M0036, 14F22M0132, 14F23B0042, 14F23B0142, 14F23B0143, 14F23B0144, 14F23B0154, 14F23D0203, 14F23D0204, 14F23D0205, 14F23D0206, 14F23D0207, 14F23E0093, 14F23E0094, 14F23E0210, 14F23E0211, 14F23F0036, 14F23F0037, 14F23F0038, 14F23F0252, 14F23F0253, 14F23F0377, 14F23F0378, 14F23F0379, 14F23G0147, 14F23G0148, 14F23G0149, 14F23G0150, 14F23G0151, 14F23G0252, 14F23H0087, 14F23H0088, 14F23H0089, 14F23H0090, 14F23H0163, 14F23H0253, 14F23H0254, 14F23H0255, 14F23J0031, 14F23J0032, 14F23J0137, 14F23J0138, 14F23J0276, 14F23J0277; l) REF RA-04220-W, UDI/DI 30801902193316, Lot Numbers: 13F22E0142, 13F22H0417, 13F22J0586, 13F22M0442, 13F23C0015, 13F23J0017; UDI/DI 30801902202988, Lot Numbers: 13F22F0124, 13F23B0363, 13F23H0010, 13F23J0013, 13F23K0001

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1267-2024
openFDA file date
Oct 6, 2026
Loaded by RegDataLab
Oct 12, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it. FDA may have changed the recall since.