Z-1273-2021 Class II Terminated

Recalled by Philips Healthcare (Suzhou) Co., Ltd. — Suzhou, N/A

FDA device recall Z-1273-2021 was initiated by Philips Healthcare (Suzhou) Co., Ltd. on Feb 24, 2021 and is designated Class II. Reason for recall: Philips has internally detected an issue with the actuators used in the Incisive CT couch, which could result in the couch failure to self-lock, and unexpectedly moving slowly downward to its limit… The recall status is terminated (terminated Feb 17, 2023). Affected quantity: 109 systems.

Recall details

Product Type
Devices
Report Date
Mar 24, 2021
Initiation Date
Feb 24, 2021
Termination Date
Feb 17, 2023
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
109 systems

Product description

Philips CT systems are advanced continuous-rotation computed tomography systems suitable for a wide range of computed tomographic (CT) applications - Product Usage: intended for use in the head and whole body.

Reason for recall

Philips has internally detected an issue with the actuators used in the Incisive CT couch, which could result in the couch failure to self-lock, and unexpectedly moving slowly downward to its limit position.

Distribution pattern

Worldwide distribution - US Nationwide distribution in the states of KY, FL, St. Croix, and the countries of Australia, Bolivia, China, Germany, Hungary, Italy, Japan, Republic of Korea, Latvia, Palestine, Puerto Rico, Romania, Russia, Saudi Arabia, Senegal, Serbia, Spain, Sweden, Thailand, Uzbekistan.

Code information

Unique Device Identifier (01)00884838059849(21)304034 (01)00884838059849(21)325078 (01)00884838085015(21)500218 (01)00884838085015(21)500219 (01)00884838085015(21)500221 (01)00884838085015(21)500222 (01)00884838085015(21)500223 (01)00884838085015(21)500224 (01)00884838085015(21)500225 (01)00884838085015(21)500226 (01)00884838085015(21)500227 (01)00884838085015(21)500229 (01)00884838085015(21)500231 (01)00884838085015(21)500233 (01)00884838085015(21)500234 (01)00884838085015(21)500237 (01)00884838085015(21)500238 (01)00884838085015(21)500239 (01)00884838085015(21)500241 (01)00884838085015(21)500242 (01)00884838085015(21)500245 (01)00884838085015(21)500247 (01)00884838085015(21)500248 (01)00884838085015(21)500249 (01)00884838085015(21)500251 (01)00884838085015(21)500252 (01)00884838085015(21)500254 (01)00884838085015(21)500255 (01)00884838085015(21)500256 (01)00884838085015(21)500257 (01)00884838085015(21)500258 (01)00884838085015(21)500261 (01)00884838085015(21)500264 (01)00884838085015(21)500274 (01)00884838085015(21)500283 (01)00884838085015(21)500296 (01)00884838085015(21)530038 (01)00884838085015(21)530067 (01)00884838085015(21)530071 (01)00884838085015(21)530098 (01)00884838085015(21)530099 (01)00884838085015(21)530101 (01)00884838085015(21)530103 (01)00884838085015(21)530104 (01)00884838085015(21)530106 (01)00884838085015(21)530107 (01)00884838085015(21)530108 (01)00884838085015(21)530109 (01)00884838085015(21)530111 (01)00884838085015(21)530112 (01)00884838085015(21)530113 (01)00884838085015(21)530114 (01)00884838085015(21)530115 (01)00884838085015(21)530116 (01)00884838085015(21)530117 (01)00884838085015(21)530118 (01)00884838085015(21)530119 (01)00884838085015(21)530120 (01)00884838085015(21)530121 (01)00884838085015(21)530124 (01)00884838085015(21)530126 (01)00884838085015(21)530127 (01)00884838085015(21)530128 (01)00884838085015(21)530129 (01)00884838085015(21)530130 (01)00884838085015(21)530132 (01)00884838085015(21)530133 (01)00884838085015(21)530134 (01)00884838085015(21)530135 (01)00884838085015(21)530136 (01)00884838085015(21)530137 (01)00884838085015(21)530138 (01)00884838085015(21)530141 (01)00884838085015(21)530148 (01)00884838085015(21)530150 (01)00884838085015(21)530152 (01)00884838085015(21)530156 (01)00884838085015(21)530157 (01)00884838085015(21)530159 (01)00884838085015(21)530160 (01)00884838085015(21)530163 (01)00884838085015(21)530165 (01)00884838085015(21)530166 (01)00884838085015(21)530167 (01)00884838085015(21)530169 (01)00884838085015(21)530170 (01)00884838085015(21)530171 (01)00884838085015(21)530172 (01)00884838085015(21)530173 (01)00884838085015(21)530174 (01)00884838085015(21)530175 (01)00884838085015(21)530176 (01)00884838085015(21)530177 (01)00884838085015(21)530178 (01)00884838085015(21)530180 (01)00884838085015(21)530181 (01)00884838085015(21)530182 (01)00884838085015(21)530183 (01)00884838085015(21)530186 (01)00884838085015(21)530187 (01)00884838085015(21)530188 (01)00884838085015(21)530189 (01)00884838085015(21)530190 (01)00884838085015(21)530191 (01)00884838085015(21)530192 (01)00884838085015(21)530194 (01)00884838085015(21)530197 (01)00884838085015(21)530198 (01)00884838085015(21)500230

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1273-2021
openFDA file date
Nov 10, 2026
Loaded by RegDataLab
Nov 11, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Source: openFDA record for Z-1273-2021. Report a data issue: hello@regdatalab.com.