Z-1279-2024 Class II Ongoing
Recall Details
- Product Type
- Devices
- Report Date
- March 20, 2024
- Initiation Date
- October 6, 2023
- Termination Date
- N/A
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 10 units
Product Description
D-Clip Standard/Mini Applier, 230mm bayonet, jaw rigid, double action.
Reason for Recall
Customer complaint received that reported the applier failed during surgery. The applier did not release the clip. The applier remained attached to the clip.
Distribution Pattern
US Nationwide distribution in the state of ME.
Code Information
Model number: 35.412; UDI/DI: 04250603739616; All Lot numbers.