Z-1283-2023 Class II Ongoing

Recalled by DeRoyal Industries Inc — Powell, TN

FDA device recall Z-1283-2023 was initiated by DeRoyal Industries Inc on Feb 9, 2023 and is designated Class II. Reason for recall: The reason for the voluntary recall is due to shrinkage of the canister lid. This shrinkage causes the lid not to contact the outer shell to create a seal and thus, it does not create and hold vacuum. The recall status is ongoing. Affected quantity: 27,200 devices.

Recall details

Product Type
Devices
Report Date
Mar 29, 2023
Initiation Date
Feb 9, 2023
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
27,200 devices

Product description

DeRoyal SafeLiner Suction Canister, 1000cc, REF PHESL-1000B

Reason for recall

The reason for the voluntary recall is due to shrinkage of the canister lid. This shrinkage causes the lid not to contact the outer shell to create a seal and thus, it does not create and hold vacuum.

Distribution pattern

US Nationwide distribution in the states of Arizona, Texas, Virginia, Pennsylvania, California, Missouri, North Carolina, Illinois, Indiana, Ohio, Arkansas.

Code information

GTIN 50749756377046, Lot 58447585 exp. 12/22/2027

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1283-2023
openFDA file date
Nov 19, 2026
Loaded by RegDataLab
Nov 20, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for DeRoyal Industries Inc? Ask Claude or ChatGPT with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. The API key is free: no card.

Get a free API key

How the connector works

Source: openFDA record for Z-1283-2023. Report a data issue: hello@regdatalab.com.