Z-1289-2015 Class III Terminated
FDA device recall Z-1289-2015 was initiated by Siemens Healthcare Diagnostics, Inc. on Jan 15, 2015 and is designated Class III. Reason for recall: Complaints have been confirmed on failed proficiency testing due to serum samples recovering too low when using some N Latex IgM reagent kit lots 44018, 44019 and 44042. The recall status is terminated (terminated Aug 17, 2016). Affected quantity: 26.
Recall details
- Product Type
- Devices
- Report Date
- Mar 25, 2015
- Initiation Date
- Jan 15, 2015
- Termination Date
- Aug 17, 2016
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 26
Product description
N Latex IgM; In-vitro diagnostic reagent for the quantitative determination of IgM in human cerebrospinal fluid (CSF) and in paired CSF/serum samples by means of particle-enhanced immunonephelometry using the BN II and BN ProSpec Systems. The determination of IgM aids in the evaluation of the patient's immune system.
Reason for recall
Complaints have been confirmed on failed proficiency testing due to serum samples recovering too low when using some N Latex IgM reagent kit lots 44018, 44019 and 44042.
Distribution pattern
Worldwide Distribution: US distribution in states of: UT , CO, CA , FL , NY , OH , NC and country of : Canada.
Code information
Catalog # OQAC11, lots 44018, 44019 and 44042, expiration 4/3/2015