Z-1289-2015 Class III Terminated

Recalled by Siemens Healthcare Diagnostics, Inc. — Newark, DE

FDA device recall Z-1289-2015 was initiated by Siemens Healthcare Diagnostics, Inc. on Jan 15, 2015 and is designated Class III. Reason for recall: Complaints have been confirmed on failed proficiency testing due to serum samples recovering too low when using some N Latex IgM reagent kit lots 44018, 44019 and 44042. The recall status is terminated (terminated Aug 17, 2016). Affected quantity: 26.

Recall details

Product Type
Devices
Report Date
Mar 25, 2015
Initiation Date
Jan 15, 2015
Termination Date
Aug 17, 2016
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
26

Product description

N Latex IgM; In-vitro diagnostic reagent for the quantitative determination of IgM in human cerebrospinal fluid (CSF) and in paired CSF/serum samples by means of particle-enhanced immunonephelometry using the BN II and BN ProSpec Systems. The determination of IgM aids in the evaluation of the patient's immune system.

Reason for recall

Complaints have been confirmed on failed proficiency testing due to serum samples recovering too low when using some N Latex IgM reagent kit lots 44018, 44019 and 44042.

Distribution pattern

Worldwide Distribution: US distribution in states of: UT , CO, CA , FL , NY , OH , NC and country of : Canada.

Code information

Catalog # OQAC11, lots 44018, 44019 and 44042, expiration 4/3/2015

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1289-2015
openFDA file date
Nov 16, 2026
Loaded by RegDataLab
Nov 17, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Source: openFDA record for Z-1289-2015. Report a data issue: hello@regdatalab.com.