Z-1294-2018 Class II Ongoing

Recalled by Seiler Instr. & Mfg. Co. Inc. — Saint Louis, MO

FDA device recall Z-1294-2018 was initiated by Seiler Instr. & Mfg. Co. Inc. on Jan 1, 2017 and is designated Class II. Reason for recall: The microscope may tip over if attachments are fully extended in one line and there is an attempt to move the device not by the handles mounted on the post. The recall status is ongoing. Affected quantity: 43 units, but 1 affected.

Recall details

Product Type
Devices
Report Date
Apr 11, 2018
Initiation Date
Jan 1, 2017
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
43 units, but 1 affected

Product description

Alpha Air 6 ENT Microscope Microscope examination of the human ear, nose, and throat.

Reason for recall

The microscope may tip over if attachments are fully extended in one line and there is an attempt to move the device not by the handles mounted on the post.

Distribution pattern

Worldwide Distribution - US Distribution to the states of : CA, CO, FL, IL, MN, NV, OK, TX., and to the countries of : Canada, China, Ecuador, England, Greece, Guatemala, Jordan, Lebanon, Mexico, Romania, Russia, South Korea, Taiwan, Japan and United Arab Emirates.

Code information

Model Number 6 ENT

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1294-2018
openFDA file date
Nov 27, 2026
Loaded by RegDataLab
Nov 28, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Seiler Instr. & Mfg. Co. Inc.? Ask Claude with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. Free to start, no card.

Question to ask in Claude

Use RegDataLab to give me an FDA risk summary for Seiler Instr. & Mfg. Co. Inc.: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.

Ask Claude about Seiler Instr. & Mfg. Co. Inc. →

Source: openFDA record for Z-1294-2018. Report a data issue: hello@regdatalab.com.