Z-1305-2018 Class II Terminated

Recalled by Dynavox Systems Llc — Pittsburgh, PA

FDA device recall Z-1305-2018 was initiated by Dynavox Systems Llc on Dec 14, 2017 and is designated Class II. Reason for recall: The battery is integral to the device. If the device sits unused and is not charged for a lengthy period of time (a few months) the battery will go into a deep discharge state and the battery will… The recall status is terminated (terminated Dec 11, 2019). Affected quantity: 307.

Recall details

Product Type
Devices
Report Date
Apr 11, 2018
Initiation Date
Dec 14, 2017
Termination Date
Dec 11, 2019
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
307

Product description

Dynavox T15 Speech Generating Device ( including Batteries)

Reason for recall

The battery is integral to the device. If the device sits unused and is not charged for a lengthy period of time (a few months) the battery will go into a deep discharge state and the battery will expire. The battery may out-gas which may be described as swelled or puffed, as gas is trapped inside the pack surrounding the cells. This battery may cause the device case to separate a small amount.

Distribution pattern

Nationally

Code information

TL1405 TL1407 Part Numbers: T15: 750423 Date Codes are part of the device s unique Serial Number and represent the year and month the device was manufactured.

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1305-2018
openFDA file date
Nov 24, 2026
Loaded by RegDataLab
Nov 25, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Dynavox Systems Llc? Ask Claude or ChatGPT with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. The API key is free: no card.

Get a free API key

How the connector works