Z-1310-2026 Class II Ongoing

Recalled by ETAC A/S — Gedved, N/A

FDA device recall Z-1310-2026 was initiated by ETAC A/S on Nov 17, 2025 and is designated Class II. Reason for recall: A customer experienced that during assembly and load testing of the device, the bolt for mounting the lifting bar broke. Upon further investigation of this customer complaint, it became clear that the… The recall status is ongoing. Affected quantity: 64 units.

Recall details

Product Type
Devices
Report Date
Feb 11, 2026
Initiation Date
Nov 17, 2025
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
64 units

Product description

Brand Name: Molift Product Name: Molift 2-point sling bars (Accessory to mobile hoists) Model/Catalog Numbers: (1) 1530101, (2) 1530102 Product Description: The mobile hoist is an assistive device intended for alleviation of, or compensation for, a functional, impairment due to an injury or disability. The device is designed for an individual lacking the ability to stand up and transfer themselves to / from a bed, the floor, a wheelchair, a chair, a toilet or similar due to reduced mobility or physical strength.

Reason for recall

A customer experienced that during assembly and load testing of the device, the bolt for mounting the lifting bar broke. Upon further investigation of this customer complaint, it became clear that the defect occurs as a result of deformation of the materials involved between the lifting bar and the arm of the device. Such deformation can lead to limited mobility in the joint, which prevents proper functionality of the equipment.

Distribution pattern

Worldwide - US Nationwide distribution in the state of PA and the countries of Australia, Czech Republic, Finland, France, Germany, Israel, Italy, Japan, New Zealand, Norway, Poland, Qatar Spain, Sweden, Switzerland, UK.

Code information

(1) Model/Catalog Number: 1530101; UDI-DI: TBD; (2) Model/Catalog Number: 1530102; UDI-DI: TBD;

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1310-2026
openFDA file date
Nov 10, 2026
Loaded by RegDataLab
Nov 11, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Source: openFDA record for Z-1310-2026. Report a data issue: hello@regdatalab.com.