Z-1321-2020 Class II Terminated
Recall Details
- Product Type
- Devices
- Report Date
- February 26, 2020
- Initiation Date
- March 14, 2019
- Termination Date
- February 15, 2024
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 11 units
Product Description
Implant Retrieval Instrument Tri Channel 6.0 22 mm - Product Usage: They can be used to remove implants, if the implant connection has been damaged. It can be used to remove implants with internal conical connection, internal tri-channel connection or external hex connection.
Reason for Recall
Due to a problem in the packaging sealing process at the supplier of the affected devices, it cannot be guaranteed that the sterilized medical devices in the scope of this recall remain sterile during their shelf-life.
Distribution Pattern
Worldwide distribution - US Nationwide distribution and countries of CA, AU, JP, BR, ZA, RU, TW, HK, IN, CN, SA, NZ, MO, CO, LB, AT, BE, CH, CY, DE, DK, EE, ES, FI, FR, GB, GG, GR, HR, HU, IE, LT, LU, NL, PL, PT, RS, SE.
Code Information
(01)07332747083811(10)73313 Catalog# 37476