Z-1323-2019 Class II Ongoing

Recalled by Philips North America, LLC — Andover, MA

FDA device recall Z-1323-2019 was initiated by Philips North America, LLC on Apr 19, 2019 and is designated Class II. Reason for recall: System software inhibits the monitor and as a result does not measure, display and alarm for pulse rates above 240 beats per minute. The recall status is ongoing. Affected quantity: 61522.

Recall details

Product Type
Devices
Report Date
May 22, 2019
Initiation Date
Apr 19, 2019
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
61522

Product description

SureSigns VS4 NBP, SpO2, Product Number 863283. For monitoring the physiological parameters of patients (NBP, SpO2 & Temperature).

Reason for recall

System software inhibits the monitor and as a result does not measure, display and alarm for pulse rates above 240 beats per minute.

Distribution pattern

Worldwide distribution. US Nationwide including Puerto Rico, Andorra, Argentina, Aruba, Australia, Austria, Bahrain, Bangladesh, Belgium, Brazil, Brunei Darussal, Bulgaria, Canada, Chile, China, Colombia, Cyprus, Dominican Republic, Egypt, France, French Guiana, French Polynesia, Germany, Gibraltar, Guadeloupe, Hong Kong, Hungary, Iceland, India, Indonesia, Iran, Ireland, Israel, Italy, Japan, Jordan, Kenya, Korea Republic, Kosovo, Kuwait, Latvia, Lebanon, Lesotho, Libya, Luxembourg, Malaysia, Maldives, Mali, Malta, Martinique, Mexico, Morocco, Mozambique, Myanmar, Namibia, Netherlands, New Caledonia, New Zealand, Norway, Oman, Pakistan, Palestine, Philippines, Portugal, Qatar, Reunion, Romania, Rwanda, Saudi Arabia, Singapore, Slovenia, South Africa, South Korea, Spain, Sri Lanka, Switzerland, Taiwan, Tanzania, Thailand, Tunisia, Turkey, United Kingdom, United Arab Emirates,Venezuela, Vietnam, and Zimbabwe.

Code information

Serial Numbers: All Including UDIs: (01)00884838028890(21) and (01)00884838087075(21) followed by serial number.

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1323-2019
openFDA file date
Nov 27, 2026
Loaded by RegDataLab
Nov 28, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Philips North America, LLC? Ask Claude with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. Free to start, no card.

Question to ask in Claude

Use RegDataLab to give me an FDA risk summary for Philips North America, LLC: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.

Ask Claude about Philips North America, LLC →

Source: openFDA record for Z-1323-2019. Report a data issue: hello@regdatalab.com.