Z-1327-2019 Class II Terminated

Recalled by PENTAX of America Inc — Redwood City, CA

Recall Details

Product Type
Devices
Report Date
May 22, 2019
Initiation Date
December 3, 2018
Termination Date
August 14, 2020
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
4 devices

Product Description

Pentax Medical C2 CryoBalloon Standard 90 degree Catheter, Model FG 1030, Sterile, RX. Product Usage: The C2 CryoBalloon Ablation System is intended to be used as a cryosurgical tool in the field of general surgery, specifically for endoscopic application and the ablation of dysplastic Barrett s Esophagus. The system includes the following components: (1) CryoBalloon Catheter connects to the Controller, which controls the operation of the Catheter such as diffuser (sprayer) positioning and ablation (nitrous oxide release); (2) CryoBalloon Controller contains the cartridge heater and cryogen delivery valve, which is controlled with the foot pedal; (3) Foot Pedal; and (4) CryoBalloon Cartridge containing 36 grams of nitrous oxide.

Reason for Recall

Incorrect default dose and dose increments may be transmitted to the controller if the catheter RFID tag is not correct or is not being correctly or completely readable by the controller.

Distribution Pattern

Distribution was made to CA, DC, FL, LA, MA, MD, MN, NC, NJ, NY, OH, PA, SC, TN, TX, VA, and WA. There was no government/military distribution. Foreign distribution was made to Canada, Germany, and the Netherlands.

Code Information

Lot/Serial number 08062018-03