Z-1332-2021 Class I Terminated

Recalled by Tenacore LLC — Santa Ana, CA

FDA device recall Z-1332-2021 was initiated by Tenacore LLC on Feb 25, 2021 and is designated Class I. Reason for recall: Bezel repair parts used, not by the original manufacturer, to service and repair infusion pump modules, may over time experience cracking or separation of the posts from the bezel. The separation of… The recall status is terminated (terminated Aug 5, 2024). Affected quantity: 1494.

Recall details

Product Type
Devices
Report Date
Apr 14, 2021
Initiation Date
Feb 25, 2021
Termination Date
Aug 5, 2024
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
1494

Product description

Alaris Infusion Pump 8100 Bezel (plastic piece only), Part: TIPA-8100-4410

Reason for recall

Bezel repair parts used, not by the original manufacturer, to service and repair infusion pump modules, may over time experience cracking or separation of the posts from the bezel. The separation of one or more bezel repair part posts may result in free flow, over infusion, under infusion or interruption of infusion.

Distribution pattern

US distribution to states of: WA, AZ, NY, TX, LA, CA, NV, ME, MN, DE, NC, FL, MO, KY, TN, and PA

Code information

Parts distributed between July 10, 2020 and January 22, 2021. Time stamp at bezel rear side shows 5 or 6 SO Numbers: 271111, 263873, 264442, 264890, 268606, 269930, 270760, 270936, 271211, 272860, 267603, 268936, 269258, 269291, 272496, 272225, 271408, 267392

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1332-2021
openFDA file date
Nov 27, 2026
Loaded by RegDataLab
Nov 28, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Tenacore LLC? Ask Claude with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. Free to start, no card.

Question to ask in Claude

Use RegDataLab to give me an FDA risk summary for Tenacore LLC: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.

Ask Claude about Tenacore LLC →

Source: openFDA record for Z-1332-2021. Report a data issue: hello@regdatalab.com.