Z-1332-2021 Class I Terminated

Recalled by Tenacore LLC — Santa Ana, CA

Recall Details

Product Type
Devices
Report Date
April 14, 2021
Initiation Date
February 25, 2021
Termination Date
August 5, 2024
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
1494

Product Description

Alaris Infusion Pump 8100 Bezel (plastic piece only), Part: TIPA-8100-4410

Reason for Recall

Bezel repair parts used, not by the original manufacturer, to service and repair infusion pump modules, may over time experience cracking or separation of the posts from the bezel. The separation of one or more bezel repair part posts may result in free flow, over infusion, under infusion or interruption of infusion.

Distribution Pattern

US distribution to states of: WA, AZ, NY, TX, LA, CA, NV, ME, MN, DE, NC, FL, MO, KY, TN, and PA

Code Information

Parts distributed between July 10, 2020 and January 22, 2021. Time stamp at bezel rear side shows 5 or 6 SO Numbers: 271111, 263873, 264442, 264890, 268606, 269930, 270760, 270936, 271211, 272860, 267603, 268936, 269258, 269291, 272496, 272225, 271408, 267392