Z-1341-2019 Class II Terminated

Recalled by Siemens Medical Solutions USA, Inc — Malvern, PA

FDA device recall Z-1341-2019 was initiated by Siemens Medical Solutions USA, Inc on Apr 9, 2019 and is designated Class II. Reason for recall: In affected Artis systems the movement of the floating tabletop may be blocked after a collision sensor has been activated during system movement. The recall status is terminated (terminated Apr 23, 2020). Affected quantity: 459 (116 US).

Recall details

Product Type
Devices
Report Date
May 22, 2019
Initiation Date
Apr 9, 2019
Termination Date
Apr 23, 2020
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
459 (116 US)

Product description

Artis Q (material 10848280, 10848281, 10848282) - Product Usage: AXIOM Artis zee and Q/ Q.zen is a family of dedicated angiography systems developed for single and biplane diagnostic imaging and interventional procedures including, but not limited to, pediatric and obese patients. Procedures that can be performed with the AXIOM Artis zee and Q/ Q.zen family include cardiac angiography, neuro angiography, general angiography, rotational angiography, operating room angiography, multipurpose angiography and whole body radiographic/fluoroscopic procedures. AXIOM Artis zee and Q/ Q.zen can also support the acquisition of position triggered imaging for spatial data synthesis.

Reason for recall

In affected Artis systems the movement of the floating tabletop may be blocked after a collision sensor has been activated during system movement.

Distribution pattern

Worldwide distribution - US Nationwide distribution in the states of AL,AR, AZ, CA, FL, ID, IL, IN, KS, LA, MA, MD, MI, MN, MO, NC, ND, NE, NV, NY, OH, OK, OR, PA, RI, SC, SD, TX, UT, VA , WA WI and WV. and Foreign countries.

Code information

Serial Number 103302 103279 103285 103286 103277 103295 103310 103293 109569 109502 109592 109523 109583 109512 109496 109542 109589 109552 109598 109500 109504 109505 109526 109565 109536 109520 109501 109507 109567 109541 109515 109551 109518 109538 109531 109532 109543 109503 109497 109527 109560 109564 109558 109524 109525 109591 121471 121467 121501 121445 121459 121490 121461 121491 121455 121504 121457 121460 121493 121487 121472 121478 121479 121470 121485 121486 121462 121468 121505 121506 121502

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1341-2019
openFDA file date
Nov 27, 2026
Loaded by RegDataLab
Nov 28, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Siemens Medical Solutions USA, Inc? Ask Claude with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. Free to start, no card.

Question to ask in Claude

Use RegDataLab to give me an FDA risk summary for Siemens Medical Solutions USA, Inc: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.

Ask Claude about Siemens Medical Solutions USA, Inc →

Source: openFDA record for Z-1341-2019. Report a data issue: hello@regdatalab.com.