Z-1343-2016 Class II Terminated

Recalled by AngioDynamics Inc. — Glens Falls, NY

FDA device recall Z-1343-2016 was initiated by AngioDynamics Inc. on Nov 3, 2015 and is designated Class II. Reason for recall: Procedure Products initiated the recall due to the inner lumen of the needles not meeting the intended specification. Specifically, at least a portion of the affected lot is unable to accept a 0.038"… The recall status is terminated (terminated Mar 16, 2017). Affected quantity: 136 kits.

Recall details

Product Type
Devices
Report Date
Apr 13, 2016
Initiation Date
Nov 3, 2015
Termination Date
Mar 16, 2017
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
136 kits

Product description

NAMIC CONVENIENCE KIT SELF REGIONAL HEALTHCARE PV LEG PACK, REF Catalog No. 651919311, UPN Product No. H7496519193111 Product Usage - This needle is intended to be placed over the guidewire to enlarge the opening in the vessel, and is then removed before sliding the catheter over the guidewire

Reason for recall

Procedure Products initiated the recall due to the inner lumen of the needles not meeting the intended specification. Specifically, at least a portion of the affected lot is unable to accept a 0.038" guidewire as intended. In some cases the inner lumen was found to be 0.036".

Distribution pattern

US Nationwide in the states of CA, IL, MA, SC, TX

Code information

Batch/Lot No. 4921353 Use By 2018-05 and Batch/Lot No. 4932710 Use By 2018-06-30

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1343-2016
openFDA file date
Nov 24, 2026
Loaded by RegDataLab
Nov 25, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for AngioDynamics Inc.? Ask Claude or ChatGPT with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. The API key is free: no card.

Get a free API key

How the connector works