Z-1345-2013 Class II Terminated

Recalled by Technidata S.A. — Montbonnot St Martin, N/A

FDA device recall Z-1345-2013 was initiated by Technidata S.A. on Jul 29, 2010 and is designated Class II. Reason for recall: Technidata has discovered that using the F8 key in the ERM session to add a combined test can result in the modification of the value of a result that has already been entered for a test of the… The recall status is terminated (terminated Dec 5, 2013). Affected quantity: 10.

Recall details

Product Type
Devices
Report Date
May 29, 2013
Initiation Date
Jul 29, 2010
Termination Date
Dec 5, 2013
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
10

Product description

TD-Synergy. Laboratory Information System.

Reason for recall

Technidata has discovered that using the F8 key in the ERM session to add a combined test can result in the modification of the value of a result that has already been entered for a test of the request.

Distribution pattern

Worldwide Distribution-USA (nationwide) including Washington DC and the states of California, Nevada, Kentucky, Alabama, Rhode Island, Pennsylvania, and Louisiana and the country of Italy.

Code information

TD Lims product versions: LIS (ALLX)-21.041-_ML-5370, LIS (ALLX)-22.021-_ML-5374, TD-Synergy product versions: SYN (ALLX)- 03.13B-_ML-5368, SYN (ALLX)- 03.141-_ML-5369, SYN(LINUX)-06.011-_ML-5381, SYN (ALLX)-11.11B-_ML-5345, SYN (ALLX)-11.21B-_ML-5346 SYN (ALL)- 11.31C-_ML-5372, SYN (LINUX)-11.321-_FR-5382, SYN (ALL)-11.321-_ML-5373

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1345-2013
openFDA file date
Nov 16, 2026
Loaded by RegDataLab
Nov 17, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Source: openFDA record for Z-1345-2013. Report a data issue: hello@regdatalab.com.