Z-1351-2019 Class II Ongoing
Recall Details
- Product Type
- Devices
- Report Date
- May 22, 2019
- Initiation Date
- February 4, 2019
- Termination Date
- N/A
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 427 total
Product Description
Molift 2-Point sling bar, an accessory for Molift Mover 205. Item nos. 1430016 1430017 1430021
Reason for Recall
The firm has received 7 reports of malfunction for the device in which the hooks have broken off, potentially leading the sling to slip out of the sling bar and causing the patient to fall.
Distribution Pattern
Domestic distribution to Arizona, Delaware, Florida, Georgia, Massachusetts, Ohio, South Carolina, Texas, Washington, and Wisconsin. Foreign distribution to Canada, Asia, Europe, and Australia.
Code Information
All units manufactured between May 2013 and October 2018. Date of manufacture is labeled on the product.