Z-1365-2020 Class II Terminated

Recalled by BIOMERIEUX - Grenoble — Grenoble Cedex 1, N/A

FDA device recall Z-1365-2020 was initiated by BIOMERIEUX - Grenoble on Jan 10, 2020 and is designated Class II. Reason for recall: Following a customer complaint about colored eluates, leading to invalid results bioMerieux has confirmed an issue with NUCLISENS Lysis Buffer, ref. 200292, lot 19040201. The issue is linked to drift… The recall status is terminated (terminated Apr 14, 2022). Affected quantity: 1024.

Recall details

Product Type
Devices
Report Date
Mar 4, 2020
Initiation Date
Jan 10, 2020
Termination Date
Apr 14, 2022
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
1024

Product description

NUCLISENS¿ Lysis Buffer, IVD, REF 200292, 48X2 ML LYS Manufacturer bioMerieux SA 376, Chemin de l'Orme 69280 Marcy-l'Etoile - France - Product Usage: to be used for the release of total nucleic acid from biological specimens.

Reason for recall

Following a customer complaint about colored eluates, leading to invalid results bioMerieux has confirmed an issue with NUCLISENS Lysis Buffer, ref. 200292, lot 19040201. The issue is linked to drift of pH electrode dysfunction.

Distribution pattern

No US distribution. No Mexican consignees. International distribution to South Africa, Uruguay, Tanzania, Romania, Netherlands, Macao, Italy, India, Hong Kong, United Kingdom, France, Spain, Germany, China, Chile, Canada, Belgium, Australia, Argentina, Curacao.

Code information

Lot 19040201, Expiry 28-Feb-2021

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1365-2020
openFDA file date
Nov 27, 2026
Loaded by RegDataLab
Nov 28, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for BIOMERIEUX - Grenoble? Ask Claude with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. Free to start, no card.

Question to ask in Claude

Use RegDataLab to give me an FDA risk summary for BIOMERIEUX - Grenoble: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.

Ask Claude about BIOMERIEUX - Grenoble →

Source: openFDA record for Z-1365-2020. Report a data issue: hello@regdatalab.com.