Z-1370-2015 Class II Terminated
FDA device recall Z-1370-2015 was initiated by BD Biosciences, Systems & Reagents on Mar 6, 2015 and is designated Class II. Reason for recall: The circuit breakers on some BD FACSCalibur instruments may have the incorrect rated amperage (20 amps instead of 2 amps) causing the circuit breaker to fail to open in an over-current situation. The recall status is terminated (terminated Apr 4, 2016). Affected quantity: 72 devices.
Recall details
- Product Type
- Devices
- Report Date
- Apr 8, 2015
- Initiation Date
- Mar 6, 2015
- Termination Date
- Apr 4, 2016
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 72 devices
Product description
Reason for recall
Distribution pattern
Worldwide Distribution -US including TX, NJ, and OK and Internationally to Hong Kong, Japan, Singopore, and Belguim