Z-1370-2015 Class II Terminated

Recalled by BD Biosciences, Systems & Reagents — San Jose, CA

FDA device recall Z-1370-2015 was initiated by BD Biosciences, Systems & Reagents on Mar 6, 2015 and is designated Class II. Reason for recall: The circuit breakers on some BD FACSCalibur instruments may have the incorrect rated amperage (20 amps instead of 2 amps) causing the circuit breaker to fail to open in an over-current situation. The recall status is terminated (terminated Apr 4, 2016). Affected quantity: 72 devices.

Recall details

Product Type
Devices
Report Date
Apr 8, 2015
Initiation Date
Mar 6, 2015
Termination Date
Apr 4, 2016
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
72 devices

Product description

BD FACSCalibur; Catalog Numbers 342973, 342975, 34976 In vitro Diagnostics Use.

Reason for recall

The circuit breakers on some BD FACSCalibur instruments may have the incorrect rated amperage (20 amps instead of 2 amps) causing the circuit breaker to fail to open in an over-current situation.

Distribution pattern

Worldwide Distribution -US including TX, NJ, and OK and Internationally to Hong Kong, Japan, Singopore, and Belguim

Code information

Serial Nos. E34297300656, E34297300657, E34297300659, E34297300660, E34297300661, E34297300666, E34297300667, E34297300668, E34297300669, E34297300670, E34297300671, E34297300673, E34297300675, E34297300676, E34297300677, E34297300678 E34297502563, E34297502564, E34297502565, E34297502566, E34297502567, E34297502568, E34297502569, E34297502570, E34297502571, E34297502572, E34297502573, E34297502574, E34297502575, E34297502576, E34297502577, E34297502578, E34297502579, E34297502580, E34297502581, E34297502583, E34297502584, E34297502585, E34297502586, E34297502587, E34297502588, E34297502589, E34297502590, E34297502591, E34297502592, E34297502593, E34297502594, E34297502595, E34297502596, E34297502597, E34297502599, E34297502600, E34297502601, E34297502602, E34297502603, E34297502604, E34297502605, E34297502606, E34297502607, E34297502608, E34297502612, E34297502613, E34297502619, E34297502620, E34297502621, E34297502629, E34297600499, E34297600501, E34297600502, E34297600503, E34297600507, E34297600508.

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1370-2015
openFDA file date
Nov 27, 2026
Loaded by RegDataLab
Nov 28, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for BD Biosciences, Systems & Reagents? Ask Claude with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. Free to start, no card.

Question to ask in Claude

Use RegDataLab to give me an FDA risk summary for BD Biosciences, Systems & Reagents: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.

Ask Claude about BD Biosciences, Systems & Reagents →

Source: openFDA record for Z-1370-2015. Report a data issue: hello@regdatalab.com.