Z-1370-2015 Class II Terminated
Recall Details
- Product Type
- Devices
- Report Date
- April 8, 2015
- Initiation Date
- March 6, 2015
- Termination Date
- April 4, 2016
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 72 devices
Product Description
BD FACSCalibur; Catalog Numbers 342973, 342975, 34976 In vitro Diagnostics Use.
Reason for Recall
The circuit breakers on some BD FACSCalibur instruments may have the incorrect rated amperage (20 amps instead of 2 amps) causing the circuit breaker to fail to open in an over-current situation.
Distribution Pattern
Worldwide Distribution -US including TX, NJ, and OK and Internationally to Hong Kong, Japan, Singopore, and Belguim
Code Information
Serial Nos. E34297300656, E34297300657, E34297300659, E34297300660, E34297300661, E34297300666, E34297300667, E34297300668, E34297300669, E34297300670, E34297300671, E34297300673, E34297300675, E34297300676, E34297300677, E34297300678 E34297502563, E34297502564, E34297502565, E34297502566, E34297502567, E34297502568, E34297502569, E34297502570, E34297502571, E34297502572, E34297502573, E34297502574, E34297502575, E34297502576, E34297502577, E34297502578, E34297502579, E34297502580, E34297502581, E34297502583, E34297502584, E34297502585, E34297502586, E34297502587, E34297502588, E34297502589, E34297502590, E34297502591, E34297502592, E34297502593, E34297502594, E34297502595, E34297502596, E34297502597, E34297502599, E34297502600, E34297502601, E34297502602, E34297502603, E34297502604, E34297502605, E34297502606, E34297502607, E34297502608, E34297502612, E34297502613, E34297502619, E34297502620, E34297502621, E34297502629, E34297600499, E34297600501, E34297600502, E34297600503, E34297600507, E34297600508.