Z-1396-2026 Class II Ongoing

Recalled by Medline Industries, LP — Northfield, IL

Recall Details

Product Type
Devices
Report Date
February 25, 2026
Initiation Date
January 7, 2026
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
23238 units

Product Description

Medline Convenience Kits: 1) BAPTIST FLOYD ENDO GI KIT, Model Number: DYKE1455B; 2) GI LAB OTHER ENDO KIT, Model Number: DYKE1721

Reason for Recall

Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products.

Distribution Pattern

Worldwide distribution - US Nationwide.

Code Information

1) DYKE1455B, UDI-DI: 10193489760781(each), 40193489760782(case), Lot Number: 22IBV801; 2) DYKE1721, UDI-DI: 10193489258684(each), 40193489258685(case), Lot Number: 23KMC417; 3) DYKE1721, UDI-DI: 10193489258684(each), 40193489258685(case), Lot Number: 23IMF495; 4) DYKE1721, UDI-DI: 10193489258684(each), 40193489258685(case), Lot Number: 23HMG566; 5) DYKE1721, UDI-DI: 10193489258684(each), 40193489258685(case), Lot Number: 23GMF416; 6) DYKE1721, UDI-DI: 10193489258684(each), 40193489258685(case), Lot Number: 23GMC067; 7) DYKE1721, UDI-DI: 10193489258684(each), 40193489258685(case), Lot Number: 23EMF635; 8) DYKE1721, UDI-DI: 10193489258684(each), 40193489258685(case), Lot Number: 23CMG332; 9) DYKE1721, UDI-DI: 10193489258684(each), 40193489258685(case), Lot Number: 23CMC558; 10) DYKE1721, UDI-DI: 10193489258684(each), 40193489258685(case), Lot Number: 23CMB706; 11) DYKE1721, UDI-DI: 10193489258684(each), 40193489258685(case), Lot Number: 23BMC768; 12) DYKE1721, UDI-DI: 10193489258684(each), 40193489258685(case), Lot Number: 22LMF786; 13) DYKE1721, UDI-DI: 10193489258684(each), 40193489258685(case), Lot Number: 22JMC784; 14) DYKE1721, UDI-DI: 10193489258684(each), 40193489258685(case), Lot Number: 22HMD809; 15) DYKE1721, UDI-DI: 10193489258684(each), 40193489258685(case), Lot Number: 22GMH729; 16) DYKE1721, UDI-DI: 10193489258684(each), 40193489258685(case), Lot Number: 22FMH954; 17) DYKE1721, UDI-DI: 10193489258684(each), 40193489258685(case), Lot Number: 22FMD365; 18) DYKE1721, UDI-DI: 10193489258684(each), 40193489258685(case), Lot Number: 22EMI534; 19) DYKE1721, UDI-DI: 10193489258684(each), 40193489258685(case), Lot Number: 22EMH737; 20) DYKE1721, UDI-DI: 10193489258684(each), 40193489258685(case), Lot Number: 21KBP287; 21) DYKE1721, UDI-DI: 10193489258684(each), 40193489258685(case), Lot Number: 21IBL687; 22) DYKE1721, UDI-DI: 10193489258684(each), 40193489258685(case), Lot Number: 21CBH946; 23) DYKE1721, UDI-DI: 10193489258684(each), 40193489258685(case), Lot Number: 21CBH534; 24) DYKE1721, UDI-DI: 10193489258684(each), 40193489258685(case), Lot Number: 21BBK795