Z-1398-2025 Class II Ongoing

Recalled by Merge Healthcare, Inc. — Hartland, WI

Recall Details

Product Type
Devices
Report Date
March 26, 2025
Initiation Date
February 19, 2025
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
2

Product Description

VERICIS, Merge Cardio, Model/Catalog Number: Version 9.x, Software Version: 9.0.6 and 9.0.8; VERICIS is an integrated cardiovascular information system classified as a picture archiving and communications system.

Reason for Recall

When following a specific less typical workflow in which the user remeasures using the ultrasound device or manually edits primitive or derived measurement values directly in Clinical Reporting, derived measurements may be inconsistent with the associated primitive measurements in the final patient report.

Distribution Pattern

US

Code Information

The following versions were introduced prior to the implementation of UDI. Version Numbers: 9.0.6, 9.0.8, Part Numbers: 88-00061-00, 88-00062-00, 88-00063-00, 88-00054-00, 88-00064-00,