Z-1406-2017 Class II Terminated
FDA device recall Z-1406-2017 was initiated by ErgoSafe Products, LLC (DBA) Prism Medical on Jan 30, 2017 and is designated Class II. Reason for recall: During internal cycle test at max load (625 lbs.) 2 units had straps break prematurely. The recall status is terminated (terminated Dec 4, 2017). Affected quantity: 76.
Recall details
- Product Type
- Devices
- Report Date
- Mar 15, 2017
- Initiation Date
- Jan 30, 2017
- Termination Date
- Dec 4, 2017
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 76
Product description
Reason for recall
Distribution pattern
Product was distributed to MT and NV; one VA consignee.
Code information
Source
- Dataset
- openFDA device recall enforcement reports
- FDA record
- This recall in openFDA (JSON)
- Enforcement Report
- Search FDA's Enforcement Report for Z-1406-2017
- openFDA file date
- Nov 27, 2026
- Loaded by RegDataLab
- Nov 28, 2026
Need the full FDA record for ErgoSafe Products, LLC (DBA) Prism Medical? Ask Claude with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. Free to start, no card.
Question to ask in Claude
Use RegDataLab to give me an FDA risk summary for ErgoSafe Products, LLC (DBA) Prism Medical: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.