Z-1425-2021 Class II Ongoing
FDA device recall Z-1425-2021 was initiated by Remel Inc on Mar 10, 2021 and is designated Class II. Reason for recall: When tested with H. influenza ATCC 49247 and ATCC 49766 both organisms may exhibit lighter than normal growth making zones difficult to measure. The recall status is ongoing. Affected quantity: 19 units.
Recall details
- Product Type
- Devices
- Report Date
- Apr 21, 2021
- Initiation Date
- Mar 10, 2021
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 19 units
Product description
Remel Haemophilus Test Medium(100mm) 10/PK, REF: R01503 Remel Haemophilus Test Medium (HTM) is a solid medium recommended for use in qualitative procedures by the Clinical and Laboratory Standards Institute (CLSI) as the growth medium for antimicrobial disk diffusion tests with Haemophilus species.
Reason for recall
When tested with H. influenza ATCC 49247 and ATCC 49766 both organisms may exhibit lighter than normal growth making zones difficult to measure.
Distribution pattern
US Nationwide distribution.
Code information
Lot Numbers: Lot 220711 Exp. 2021-2-17