Z-1441-2014 Class II Terminated

Recalled by Remington Medical Inc. — Alpharetta, GA

Recall Details

Product Type
Devices
Report Date
April 16, 2014
Initiation Date
March 5, 2014
Termination Date
April 9, 2018
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
41,499 cables

Product Description

FAS-LOC 6 Foot Disposable Extension Cable with Safe Connect/Screw Down. Electrical extension cable intended to transmit signal from, or power or excitation signal to patient connected electrodes. The cable is a sterile, disposable device.

Reason for Recall

Remington Medical discovered an error on the FL-601-97 IFU; part number IFU-RM-0154. Specifically, the polarity for the black chuck in the picture is wrong. It shows (+), but should show (-).

Distribution Pattern

USA (nationwide) Distribution including Canada

Code Information

Lot numbers 131901, 132262, 132263, 132533, 132693, 132694, 132833 132835, 132986, and 133171.