Z-1444-2022 Class II Terminated
Recall Details
- Product Type
- Devices
- Report Date
- July 27, 2022
- Initiation Date
- September 8, 2020
- Termination Date
- April 10, 2024
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 24
Product Description
AESCULAP S4 Cervical Cross Connector L24mm, S4C CROSS CONNECTOR FIXED 24MM, product code SWT113T. Used in spinal fusion surgery.
Reason for Recall
Insufficient clamping force of the cross connectors in the spinal surgical system which may lead to movement of the connectors on the rods and/or microparticle abrasion.
Distribution Pattern
United States Nationwide distribution in the states of CA, CO, IL, IN, MI, NC, NV, OH, PA, SC, TX & WY.
Code Information
UDI-DI: 04038653362301