Z-1446-2022 Class II Terminated

Recalled by Aesculap Implant Systems LLC — Center Valley, PA

FDA device recall Z-1446-2022 was initiated by Aesculap Implant Systems LLC on Sep 8, 2020 and is designated Class II. Reason for recall: Insufficient clamping force of the cross connectors in the spinal surgical system which may lead to movement of the connectors on the rods and/or microparticle abrasion. The recall status is terminated (terminated Apr 10, 2024). Affected quantity: 23.

Recall details

Product Type
Devices
Report Date
Jul 27, 2022
Initiation Date
Sep 8, 2020
Termination Date
Apr 10, 2024
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
23

Product description

AESCULAP S4 Cervical Cross Connector L 28-33mm, S4C CROSS CONNECTOR VARIABLE 28-33MM, SW115T. Used in spinal fusion surgery.

Reason for recall

Insufficient clamping force of the cross connectors in the spinal surgical system which may lead to movement of the connectors on the rods and/or microparticle abrasion.

Distribution pattern

United States Nationwide distribution in the states of CA, CO, IL, IN, MI, NC, NV, OH, PA, SC, TX & WY.

Code information

UDI-DI: 04038653362325

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1446-2022
openFDA file date
Nov 25, 2026
Loaded by RegDataLab
Nov 26, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.