Z-1448-2015 Class II Terminated
FDA device recall Z-1448-2015 was initiated by Vitrolife Inc on December 22, 2014 and is designated Class II. Reason for recall: The firm is recalling three batches (FP14A09, FP14A10, FP14A11) of SpermMAR IgA test due to new formulation causing the motility of sperm cells to reduce in certain samples when the new bead solution … The recall status is terminated (terminated May 5, 2015). Affected quantity: 45 kits.
Recall Details
- Product Type
- Devices
- Report Date
- April 22, 2015
- Initiation Date
- December 22, 2014
- Termination Date
- May 5, 2015
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 45 kits
Product Description
SpermMar Test 0.7ml Beads Particles Label on bottle: SpermMar Test IgA 0.7ml Beads/Particles-50 Tests IND REF SPMA_S LOT Expiration DO NOT FREEZE/NE PAS CONGELER FertiPro N.V. Label on box: SpermMar Test IGA 50 Determination Test Kit CONTENTS 1 vial SpermMar Latex Particles 0.7 ml REF SPMA_S FertiPro FertiPro N.V. Label on box: SpermMar Test IgA 50 Determination Test Kit Contents: 1 vial SpermMar latex particles, 0.7ml FertiPro N.V. For In Vitro Diagnostic Use SpermMar IqA Test Manufacturer's Product Number/Catalog Number: SPMA_S Lot/Serial Number Expiration Date FP14A09 02/29/2016 FP14A10 04/30/2016 FP14A11 04/30/2016 Product is a solution in filled into clear plastic bottles that is placed into a cardboard box. Both packages contain labeling information.
Reason for Recall
The firm is recalling three batches (FP14A09, FP14A10, FP14A11) of SpermMAR IgA test due to new formulation causing the motility of sperm cells to reduce in certain samples when the new bead solution was added.
Distribution Pattern
US Distribution to the state of : OR., and Internationally to Argentina.
Code Information
Manufacturer's Product Number/Catalog Number: SPMA_S Lot/Serial Number Expiration Date FP14A09 02/29/2016 FP14A10 04/30/2016 FP14A11 04/30/2016