Z-1449-2014 Class II Terminated

Recalled by Roche Diagnostics Operations, Inc. — Indianapolis, IN

FDA device recall Z-1449-2014 was initiated by Roche Diagnostics Operations, Inc. on Mar 13, 2014 and is designated Class II. Reason for recall: The Initial Cassette Volume Check (ICVC) is not activated on the cobas c 502 module of the cobas 8000 modular analyzer series when manually filled cobas c packs are loaded. This issue occurs with… The recall status is terminated (terminated Jan 10, 2018). Affected quantity: 76 units.

Recall Details

Product Type
Devices
Report Date
Apr 16, 2014
Initiation Date
Mar 13, 2014
Termination Date
Jan 10, 2018
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
76 units

Product Description

Cobas c8000 Modular Analyzer Series Cobas c502, Part Number 5964067001, a clinical chemistry analyzer intended for the in vitro quantitative / qualitative determination of analytes in body fluids.

Reason for Recall

The Initial Cassette Volume Check (ICVC) is not activated on the cobas c 502 module of the cobas 8000 modular analyzer series when manually filled cobas c packs are loaded. This issue occurs with tests that use cobas c pack MULTI or empty pre-labeled cobas c packs. The ICVC feature allows the reagent probe to dive into the reagent cassette and confirm the reagent level in the bottles. This ensures

Distribution Pattern

USA Nationwide Distribution including San Juan PR

Code Information

not applicable

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1449-2014
openFDA file date
Oct 23, 2026
Loaded by RegDataLab
Oct 26, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.