Z-1459-2020 Class II Terminated

Recalled by Braemar Manufacturing, LLC — Eagan, MN

FDA device recall Z-1459-2020 was initiated by Braemar Manufacturing, LLC on Jan 22, 2020 and is designated Class II. Reason for recall: An issue with the DL900 and DL950 Series Holter Recorder was recognized on January 1, 2020. This issue can prevent the recorder from operating normally. Braemar confirmed that beginning on January… The recall status is terminated (terminated May 27, 2022). Affected quantity: 10582.

Recall details

Product Type
Devices
Report Date
Mar 18, 2020
Initiation Date
Jan 22, 2020
Termination Date
May 27, 2022
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
10582

Product description

Braemar Manufacturing LLC DL900 Holter Monitor 1285 Corporate Center Dr. Suite 150 Eagan, MN 55121 USA Rx Only

Reason for recall

An issue with the DL900 and DL950 Series Holter Recorder was recognized on January 1, 2020. This issue can prevent the recorder from operating normally. Braemar confirmed that beginning on January 1, 2020 if a AAA battery is inserted in the recorder and a user attempts to start it, the recorder will display Error: 602 and fail to function for a new patient study. There are no actions that a clinical user can take to clear this error. Even if this error code is displayed, existing ECG recordings stored on the device can still be downloaded via the standard Holter Application Software. Error 602 may be able to be cleared by technical or engineering staff at the customer site.

Distribution pattern

Nationwide. No governmental consignees or Canadian consignees. International distribution to Germany, Spain, Switzerland, France, Finland, Vietnam, Israel, France, Hong Kong, Czech Republic, Finland, Belgium, Sweden, Mexico, Lebanon, United Arab Emirates, Bolivia, Italy, and Belgium.

Code information

All devices manufactured or distributed prior to 01/01/2020 are affected.

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1459-2020
openFDA file date
Nov 24, 2026
Loaded by RegDataLab
Nov 25, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Braemar Manufacturing, LLC? Ask Claude or ChatGPT with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. The API key is free: no card.

Get a free API key

How the connector works

Source: openFDA record for Z-1459-2020. Report a data issue: hello@regdatalab.com.