Z-1472-2026 Class II Ongoing

Recalled by Agiliti Health - Ellis — Ellis, KS

FDA device recall Z-1472-2026 was initiated by Agiliti Health - Ellis on Jan 21, 2026 and is designated Class II. Reason for recall: Agiliti identified an issue where the Adapt Pump (61600200-Adapt Pump) supplied with all Adapt Line support surfaces does not re-engage the Microclimate Management function after using the Autofirm… The recall status is ongoing. Affected quantity: 4286.

Recall details

Product Type
Devices
Report Date
Mar 4, 2026
Initiation Date
Jan 21, 2026
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
4286

Product description

Adapt Pump, Model/Catalog Number: 61600200-Adapt Pump, component of Adapt support surfaces (Adapt Air, Adapt Air EX, Adapt Air Pro, Adapt Air Pro EX, Adapt Convertible, Adapt Convertible EX), sofware version 1.3.0; Mattress, Air Flotation, Alternating Pressure

Reason for recall

Agiliti identified an issue where the Adapt Pump (61600200-Adapt Pump) supplied with all Adapt Line support surfaces does not re-engage the Microclimate Management function after using the Autofirm function for 10 minutes. This issue only occurs when the 10-minute timer is allowed to elapse after the Autofirm button is pressed after initial setup, and it only affects the Microclimate Management function.

Distribution pattern

US Nationwide distribution.

Code information

UDI-DI 0084569904914; All serial numbers with a manufacturing date prior to 12/04/2025

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1472-2026
openFDA file date
Nov 16, 2026
Loaded by RegDataLab
Nov 17, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Source: openFDA record for Z-1472-2026. Report a data issue: hello@regdatalab.com.