Z-1478-2022 Class II Ongoing

Recalled by Stryker Instruments — Portage, MI

FDA device recall Z-1478-2022 was initiated by Stryker Instruments on Jun 24, 2022 and is designated Class II. Reason for recall: There is a potential for the core diameter to be undersized, which may lead to the router breaking. A break can lead to unintended metal fragments in the surgical site, resulting in the potential for… The recall status is ongoing. Affected quantity: 543.

Recall details

Product Type
Devices
Report Date
Aug 3, 2022
Initiation Date
Jun 24, 2022
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
543

Product description

Stryker 2.3mm Tapered Router, Catalog no. 5820-071-023 instrument used in craniotomy procedures.

Reason for recall

There is a potential for the core diameter to be undersized, which may lead to the router breaking. A break can lead to unintended metal fragments in the surgical site, resulting in the potential for surgical intervention or damage to critical neurological or vascular structures.

Distribution pattern

Worldwide distribution - US Nationwide distribution in the states of Arizona, California, Florida, Georgia, Iowa, Michigan, Missouri, New Mexico, North Carolina, Ohio, Texas, Virginia and the countries of Canada, Netherlands, and Thailand.

Code information

UDI-DI (GTIN): 07613327294910 Affected Lots: 22011017 21330017 20216017 20140017 20139027

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1478-2022
openFDA file date
Nov 24, 2026
Loaded by RegDataLab
Nov 25, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Stryker Instruments? Ask Claude or ChatGPT with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. The API key is free: no card.

Get a free API key

How the connector works

Source: openFDA record for Z-1478-2022. Report a data issue: hello@regdatalab.com.