Z-1478-2024 Class II Ongoing

Recalled by Navinetics Inc — Rochester, MN

Recall Details

Product Type
Devices
Report Date
April 10, 2024
Initiation Date
February 26, 2024
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
20 kits

Product Description

NaviNetics Skull Anchor Key and Drill Kit, REF NN1215

Reason for Recall

Defective Skull Anchor Key. After attachment of the single use Skull Anchor Key to the patient, the user was unable to attach the CT Localizer to perform stereotactic Imaging. Attachment of the localizer is typically accomplished by fully screwing in the four thumbscrews of the localizer into the threaded holes found on the top surface of the Skull Anchor Key.

Distribution Pattern

US Nationwide distribution in the states of FL, MN, CO.

Code Information

UDI/DI 10850023620276, Lot Number 331401724