Z-1483-2013 Class II Terminated

Recalled by Zimmer Surgical Inc — Dover, OH

FDA device recall Z-1483-2013 was initiated by Zimmer Surgical Inc on May 7, 2013 and is designated Class II. Reason for recall: This removal is in response to complaints alleging that the Zimmer Air Dermatomes and Zimmer Air Dermatome II products would either not operate or operate intermittently. Zimmer's investigation… The recall status is terminated (terminated Dec 15, 2014). Affected quantity: 84 units.

Recall details

Product Type
Devices
Report Date
Jun 12, 2013
Initiation Date
May 7, 2013
Termination Date
Dec 15, 2014
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
84 units

Product description

The Zimmer Air Dermatome, REF 00- 8801- 001- 00 , Rx, non-sterile grafting instruments intended to provide variable graft thickness and width capabilities

Reason for recall

This removal is in response to complaints alleging that the Zimmer Air Dermatomes and Zimmer Air Dermatome II products would either not operate or operate intermittently. Zimmer's investigation determined that the planetary gear teeth were broken. This may cause a delay or postponement of the surgery or result in a suboptimal graft retrieval with potential of requiring an additional donor site.

Distribution pattern

Worldwide distribution: USA Nationwide including Puerto Rico; UAE, TAIWAN, SWEDEN, THAILAND, KOREA, SINGAPORE, HONG KONG CHINA, CHILE, CANADA, AUSTRALIA, JAPAN, ENGLAND, GERMANY, and EGYPT.

Code information

Serial Number 100211, 100338, 100515, 100560, 100647, 100755, 101104, 101345, 101562, 101769, 101890, 101983, 102560, 103407, 103509, 103816, 104529, 104863, 105049, 105072, 105604, 106394, 106474, 106624, 107861, 108103, 108203, 108210, 108339, 108911, 109445, 109543, 110034, 110837, 111046, 111047, 111048, 111049, 111050, 111051, 111052, 111053, 111054, 111055, 111056, 111057, 111058, 111059, 111060, 111061, 111062, 111063, 111064, 111065, 111066, 111067, 111068, 111069, 111071, 111072, 111073, 111074, 111075, 111076, 111079, 111081, 111083, 111084, 111085, 111086, 111087, 111089, 111090, 111091, 111092, 111096, 111100, 111101, 111102, 111104, 111105, 111116, 111121, 111122

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1483-2013
openFDA file date
Nov 27, 2026
Loaded by RegDataLab
Nov 28, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Zimmer Surgical Inc? Ask Claude with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. Free to start, no card.

Question to ask in Claude

Use RegDataLab to give me an FDA risk summary for Zimmer Surgical Inc: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.

Ask Claude about Zimmer Surgical Inc →

Source: openFDA record for Z-1483-2013. Report a data issue: hello@regdatalab.com.