Z-1496-2021 Class II Terminated

Recalled by Siemens Medical Solutions USA, Inc — Malvern, PA

FDA device recall Z-1496-2021 was initiated by Siemens Medical Solutions USA, Inc on Mar 30, 2021 and is designated Class II. Reason for recall: Due to inappropriate cleaning, some Artis systems show unexpected corrosion of visible belts which are needed to move system parts (e.g. C-Arm), Increased corrosion may lead to a malfunction of the… The recall status is terminated (terminated Jul 12, 2022). Affected quantity: 3,320 units U.S.

Recall details

Product Type
Devices
Report Date
May 5, 2021
Initiation Date
Mar 30, 2021
Termination Date
Jul 12, 2022
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
3,320 units U.S.

Product description

Artis systems - Product Usage: an angiography systems developed for single and biplane diagnostic imaging and interventional procedures including, but not limited to, pediatric and obese patients. Model Numbers: System Material # AXIOM Artis FC 05904433; AXIOM Artis FA 05904441; AXIOM Artis MP 05904466; AXIOM Artis BC 05904649; AXIOM Artis BA 05904656; AXIOM Artis dBC Mag. Nav. 05917054; AXIOM Artis TA 07007755; AXIOM Artis dTA 07008605; AXIOM Artis dFC 07412807; AXIOM Artis dTC 07413078; AXIOM Artis dBA 07555357; AXIOM Artis dMP 07555365; AXIOM Artis dFA 07555373; AXIOM Artis dFC Mag. Nav. 07727717; AXIOM Artis TC 07728350; AXIOM Artis dBC 07728392; Artis zee floor 10094135; Artis zee ceiling 10094137; Artis zee MP 10094139; Artis zee biplane 10094141; Artis zee floor MN 10094142; Artis zee biplane MN 10094143; Artis Q floor 10848280; Artis Q ceiling 10848281; Artis Q biplane 10848282; Artis Q.zen floor 10848353; Artis Q.zen ceiling 10848354; Artis Q.zen biplane 10848355.

Reason for recall

Due to inappropriate cleaning, some Artis systems show unexpected corrosion of visible belts which are needed to move system parts (e.g. C-Arm), Increased corrosion may lead to a malfunction of the belts which may result in limited functionality of the Artis system up to system failure. Unintended movement of the C-Arm may cause a hazardous situation to the patient, operator, or staff members and may result in a situation where it is necessary to cancel clinical treatment or to continue treatment on an alternative system

Distribution pattern

US Nationwide distribution.

Code information

All serial numbers

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1496-2021
openFDA file date
Nov 27, 2026
Loaded by RegDataLab
Nov 28, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Siemens Medical Solutions USA, Inc? Ask Claude with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. Free to start, no card.

Question to ask in Claude

Use RegDataLab to give me an FDA risk summary for Siemens Medical Solutions USA, Inc: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.

Ask Claude about Siemens Medical Solutions USA, Inc →

Source: openFDA record for Z-1496-2021. Report a data issue: hello@regdatalab.com.