Z-1503-2016 Class II Terminated

Recalled by Ad-Tech Medical Instrument Corporation — Racine, WI

FDA device recall Z-1503-2016 was initiated by Ad-Tech Medical Instrument Corporation on Mar 29, 2016 and is designated Class II. Reason for recall: Ad-Tech is recalling unopened/unused sterile cables intended to be used only with all compatible Ad-Tech electrodes for the purpose of connecting these electrodes to third party monitors/stimulators… The recall status is terminated (terminated Aug 24, 2016). Affected quantity: 197.

Recall details

Product Type
Devices
Report Date
Apr 27, 2016
Initiation Date
Mar 29, 2016
Termination Date
Aug 24, 2016
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
197

Product description

Ad-Tech Electrode Connection System/Cables. Lightweight TECH-ATTACH Cable Models: 1. L-DCL-4DINX (4 contacts: 1 terminal block with 4 connectors each); 2. L-DCL-6DINX (6 contacts: 1 terminal block with 6 connectors each); 3. L-DCL-8DINX (8 contacts: 1 terminal block with 8 connectors each); 4. L-DCL-16BDINX (16 contacts: 1 terminal block with 16 connectors each); 5. L-DCL-32BDINX (32 contacts: 2 terminal blocks with 16 connectors each); 6. L-DCL-64BDINX (64 contacts: 4 terminal blocks with 16 connectors each). FO TECH-ATTACH Connection Cable model FO-LDC-4DINX (4 contacts: 1 terminal block with 4 connectors each). Tech-Attach: these cables require the use of an additional component - the connector block. The tail of the electrode is inserted into the block and the block in turn mates with the cable. A key-pin is used to ensure proper orientation. The cable terminated in EEG compatible connector sockets. Lightweight CABRIO Cable, model L-SRL-64BDINX (64 contacts: 4 terminal blocks with 16 connectors each). Cabrio cables are similar to Tech-Attach cables except that the block portion is pre-attached. The tail of an electrode is inserted into this portion which then hinges shut.

Reason for recall

Ad-Tech is recalling unopened/unused sterile cables intended to be used only with all compatible Ad-Tech electrodes for the purpose of connecting these electrodes to third party monitors/stimulators. This recall has been initiated due to package failures (i.e. compromised sterile barriers) that were identified during the execution of a shipping simulation investigation.

Distribution pattern

Distributed in the states of GA and MI, and the countries of FRANCE, UNITED KINGDOM, COLOMBIA, and BRAZIL.

Code information

Lot Numbers: 208140507, 208140552, 208140556, 208140559, 208140560, 208140561, 208140562, 208140563, 208140564R, 208140579, 208140581, 208140586, 208140588R, 208140590, 208140604.

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1503-2016
openFDA file date
Nov 25, 2026
Loaded by RegDataLab
Nov 26, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Ad-Tech Medical Instrument Corporation? Ask Claude with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. Free to start, no card.

Question to ask in Claude

Use RegDataLab to give me an FDA risk summary for Ad-Tech Medical Instrument Corporation: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.

Ask Claude about Ad-Tech Medical Instrument Corporation →

Source: openFDA record for Z-1503-2016. Report a data issue: hello@regdatalab.com.