Z-1503-2017 Class II Terminated
Recall Details
- Product Type
- Devices
- Report Date
- March 29, 2017
- Initiation Date
- November 18, 2016
- Termination Date
- July 13, 2018
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 5,993 (5,080 US; 913 OUS)
Product Description
Giraffe Spot PT Ohmeda Medical Spot PT Lite Phototherapy System- The Spot PT Lite Phototherapy System provides light therapy for the treatment of hyperbilirubinemia, commonly known as neonatal jaundice, during the newborn period in the hospital
Reason for Recall
Certain power cords could overheat. The affected power cords were manufactured by Electric-Cord and are referred to as Taller Bridge power cords. The overheating of the power cord could result in fire, charring, smoke, or sparking where the power cord connects to the wall outlet which could cause thermal injury.
Distribution Pattern
Worldwide Distribution - US (nationwide and Puerto Rico) Bahamas and Internationally to UK, Trinidad & Tobago, Taiwan, Spain, Panama, Netherlands Antilles, Mexico, Lebanon, Japan, Ireland, Honduras, El Salvador, Ecuador, Costa Rica, Colombia, Canada, Brazil, and Venezuela,
Code Information
See Attached Spreadsheat