Z-1504-2013 Class II Terminated
FDA device recall Z-1504-2013 was initiated by Stryker Instruments Div. Of Stryker Corporation on May 23, 2013 and is designated Class II. Reason for recall: Based on an accelerated aging test for this product packaging, there is an indicator of potential for a breach in the sterile barrier in the header bag and potentially the blue sterilization wrap for… The recall status is terminated (terminated Apr 14, 2014). Affected quantity: 6,990 units ( 699 boxes).
Recall details
- Product Type
- Devices
- Report Date
- Jun 12, 2013
- Initiation Date
- May 23, 2013
- Termination Date
- Apr 14, 2014
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 6,990 units ( 699 boxes)
Product description
Stryker Instruments 0400-850-100 T5 zipper Toga with Peel Away, Sterile, EO; single use; Product Usage: Protection System and are intended to protect the patient, healthcare personnel and operating room personnel against contamination, exposure of infectious bodily fluids, and the transfer of microorganismsand particulate material.
Reason for recall
Based on an accelerated aging test for this product packaging, there is an indicator of potential for a breach in the sterile barrier in the header bag and potentially the blue sterilization wrap for togas within the second and third year of shelf life. The test method used involved high heat to accelerate the aging of the packaging. Stryker is conducting additional testing and assessing this pr
Distribution pattern
USA Nationwide Distribution
Code information
1006124, 1007164, 1010236, 1011296, 1103168, 1105341, and 1108634.